Site Specifications
Page 1: ...Site Specifications...
Page 2: ...site needs to be prepared according to specifications This section describes general requirements for installing the ECi ECiQ System at your laboratory including physical and environmental requirement...
Page 3: ...sical dimensions of the system The system weighs approximately 800 pounds 366 kilograms Figure 1 1 Physical Dimensions The system and printer dimensions are described below D C F E I J G A B H L M N K...
Page 4: ...83 meters 6 ft in length and detachable North America IEC 320 to NEMA 5 20R Continental Europe IEC 320 to CEE7 Shuko UPS Dimensions I 49 4 cm 19 4 in J 43 2 cm 17 0 in K 8 9 cm 3 5 in Printer Stand f...
Page 5: ...age and Frequency for printers used in North America Electrical Outlet Receptacle Type The electrical outlet receptacle type for the system is as follows North America NEMA 5 20R Continental Europe CE...
Page 6: ...cm 30 in wide Telephone Recommendations You may also find it convenient to have a telephone near the system to communicate with technical support personnel during troubleshooting sessions A dedicated...
Page 7: ...tput 4100 BTUs per hour Operating Temperature 15 30 C 59 86 F Site Relative Humidity 15 75 RH noncondensing Altitude 0 1524 2 439 km 500 8000 ft Figure 1 2 Environmental Requirements System Relocation...