Section 6 - Troubleshooting and Error Codes
ORTHO
™
Workstation for ID-MTS
™
Gel Cards Reference Guide
16
Pub. No.: J55700ENna
2013-09-04
202
Setpoint High Error:
The setpoint value read from the temperature controller’s was higher
than expected.
203
Setpoint Calculated Low Error:
The calculated temperature setpoint is less than the
acceptable minimum. This could be caused by lab environmental temperature. Let the system
stabilize for one hour, turn the instrument off then on.
204
Setpoint Calculated High Error:
The calculated temperature setpoint is higher than the
acceptable maximum. This could be caused by lab environmental temperature. Let the
system stabilize for one hour, then turn the instrument off then on.
205
Temperature Sensor 1 Low Error:
The temperature sensor value read from the temperature
controller was lower than expected.
206
Temperature Sensor 1 High Error:
The temperature sensor value read from the temperature
controller was higher than expected.
207
Heater Controller Read Error:
There was a failure reading from the temperature controller.
208
Heater Controller Write Error:
There was a failure writing to the temperature controller.
†
Test is no longer valid