Figure 20:
Confirmation after saving the spectrum
The energy-channel ratio of the RADEAGLET scales the energy E=3072 keV to channel 2048 (2k). In-
ternally, the RADEAGLET uses 4096 channels for the spectrum processing.
Changing the x-axis scaling also affects the cursor. The instrument will show the cursor position either in
channel numbers or energy units (whatever is selected).
3.3.8 Changing the y-Axis Scaling: Linear, Logarithmic or Square Root
Scintillation detectors have a certain energy dependent sensitivity. Peaks at low energies (e.g. 59 keV of
241
Am) have a higher sensitivity than peaks at the higher end of the spectrum (e.g. 1332 keV of
60
Co).
When observing this type of spectra and the y-axis has a linear scale, some peaks at higher energies
might not be visible. To see a better display of the higher energy peaks, the user may want to look at the
spectrum either using a logarithmic scale or a square root scale for the y-axis. These different scales
allow the user to visually equalize the peak heights so that a wide range of the spectra can be viewed
without zooming.
You can choose between three data representation modes by toggling between
LIN
/
LOG
/
SQRT
.
LIN
is
the normal linear mode of displaying the spectrum.
LOG
represents the logarithmic view.
SQRT
is the
square root scale for the spectrum. Figure
, p.
shows a spectrum plotted in square-root scale for
the y-axis. The bar next to the spectrum shows “SQRT” to remind you which vertical scaling is in use. In
logarithmic view, the channels with zero counts are mapped to zero.
© innoRIID GmbH
RADEAGLET User Manual• Software 2.1.16 • Document 1.32o • 2017-12-15
Page 29 of 60
Summary of Contents for RADEAGLET
Page 6: ...Page 6 of 60 RADEAGLET User Manual Software 2 1 16 Document 1 32o 2017 12 15 innoRIID GmbH...
Page 22: ...Page 22 of 60 RADEAGLET User Manual Software 2 1 16 Document 1 32o 2017 12 15 innoRIID GmbH...
Page 59: ...innoRIID GmbH RADEAGLET User Manual Software 2 1 16 Document 1 32o 2017 12 15 Page 59 of 60...