1107231N / 082020 4. DETECTIVE X APPLICATION
49
Figure 61. Optional Metadata for ID Mode Data
To add a new
Location
,
Item description
, or
Condition
, tap the associated
Edit
button to open the data entry
field. After that, recent entries for each field will be available on a drop list by tapping the down arrow next
to the field.
When entry of metadata is finished, tap
Finish
to save it and return to the Detect Mode screen.
Prior to setup, the
Reachback
button will be disabled (gray). Section 4.12 covers setup of Reachback data
transfer for ID Mode measurements.
4.7.4
Multiple ID Measurements in One Data File
For very high levels of measurement confidence, the Detective X has a
Multiple spectra analysis
option that
lets the user wrap data collected from an unknown sample with that collected from a blank sample and a
known sample, and save the data from each related measurement in the same N42.42 data file.
To configure the Detective X for this mode of operation:
On the Advanced Setup screen, tap
Analysis
to open the Analysis Settings page, then tap to mark
Multiple spectrum analysis
checkbox (see Figure 62).
In this mode, each time ID Mode is entered, the Multi ID Mode page shown in Figure 63 will be
displayed so the user can set up and annotate the multiple measurements to be made.