6.3.2.1
SYSTem:ERRor?
Function
This SYSTem command queries the error queue for an ASCII text description of the next
error and removes it from the queue. The error messages are placed in an error queue,
with a FIFO (First In-First Out) structure. This queue is summarized in the Error AVailable
(EAV) bit in the status byte.
The System Error command is extended with relevant command and protocol errors. It will
allow the user to determine:
Scheduled commands arrive too late to meet their required pre-processing time
based on their timestamps.
The user can configure protocol to flag error in situation where:
More than one position command is received during same epoch, and the
commands contradict (not complement) each other. Values are not analyzed
to determine contradiction, but only type of data (e.g., two position inform-
ation commands are deemed to contradict, even if the actual position would
not change.). In these situations only the last received information is served
(no queue system used). The default configuration is NOT to flag error.
No position information is received during the epoch. The default con-
figuration is NOT to flag error.
Command Syntax
SYSTem:ERRor[:NEXT]?
Note
All
SOURce:SCENario
commands are only available during scenario execution. If scenario
is not running these error codes are to be returned (for both set and get functions);
-191
,"Execution not in progress".
Returned format
<error number>,"<Error Description String>"
Where:
<Error Description String> = an error description as ASCII text.
6.3 Command Reference
User Manual GSG-5/6 Series Rev. 27
235
Summary of Contents for GSG-5 Series
Page 2: ......
Page 4: ...Blank page II User Manual GSG 5 6 Series...
Page 42: ...BLANK PAGE 2 5 Signal Power Level Considerations 26 User Manual GSG 5 6 Series Rev 27...