Oroboros Instruments O2k-Core Manual Download Page 7

MiPNet19.18 

B: OroboPOS service: Supplement

 

S1

 

 

 

OROBOROS INSTRUMENTS

 

 

high-resolution respirometry

 

 

Supplement A. Videosupport 

 

»

 

O2k-Videosupport

 

(online)

 

 

 

 
The following links are functional on the OROBOROS USB flash drive: 

 

 

B1_OroboPOS from Perspex housing

 

 

 

B2 Disassembly of OroboPOS

 

 

 

B3 Cathode cleaning

 

 

OroboPOS-Polishing Cloth

, 

OroboPOS-Polishing Powder

. 

 

 

B4 Anode cleaning

 

 

 

B5 Membrane mounting

 

 
 
 

Supplement B. Links to O2k-Catalogue 

 

» 

http://wiki.oroboros.at/index.php/OroboPOS

  

» 

http://wiki.oroboros.at/index.php/O2k-Catalogue:_OroboPOS_components

  

» 

http://wiki.oroboros.at/index.php/OroboPOS-Service_Kit

  

 

 

Summary of Contents for O2k-Core

Page 1: ...Videosupport Summary Service of the polarographic oxygen sensors OroboPOS POS is the basis of signal stability low noise and high time resolution Performance specifications of the OROBOROS O2k can be...

Page 2: ...ensor service are provided in the OroboPOS Service Kit In addition distilled water and 25 ammonia solution fresh are required Store the OroboPOS in the dark 2 Cleaning Prevent damage by electrostatic...

Page 3: ...or one minute in a figure eight motion Wash the polishing powder carefully off the sensor with distilled water Subsequently the anode is always cleaned as well Section 2 2 If noise remains high or the...

Page 4: ...me of 20 ml Store at room temperature or 4 C in a closed vial To prevent the formation of gas bubbles in the electrolyte heat the electrolyte solution by shaking the electrolyte vial in hot water 40 7...

Page 5: ...le connector next to the control light of the stirrer O2k Main Unit The red dot on the male plug has to face accurately upwards when inserting the plug Avoid bending and torsions of the cable MiPNet19...

Page 6: ...ody at 60 C for a few days the zero current should be reduced In addition the zero current of the bare cable without sensor head connected is tested for any leak currents If the latter test excludes a...

Page 7: ...ROBOROS USB flash drive B1_OroboPOS from Perspex housing B2 Disassembly of OroboPOS B3 Cathode cleaning OroboPOS Polishing Cloth OroboPOS Polishing Powder B4 Anode cleaning B5 Membrane mounting Supple...