MiPNet19.18
B: OroboPOS service
18
OROBOROS INSTRUMENTS
O2k – tested and trusted
Removal of a used membrane
from the OroboPOS.
2
2. Grab the PEEK
membrane ring
on the sensor
head
3
4
3. Pull the PEEK
membrane ring from
the sensor head
4. Remove and
discharge the old
membrane
Remove an old
membrane from
the OroboPOS
OroboPOS-
Mounting Tool
POS on POS
connector
1. Remove
POS seal tip
1
1.
OroboPOS-Service Kit
Accessories for sensor service are provided in the
In addition, distilled water and 25% ammonia solution (fresh) are
required. Store the OroboPOS in the dark.
2.
Cleaning
Prevent damage by electrostatic
discharge (ESD) when handling the
or
cable
connections to the O2k (
).
For sensor service, remove the black
small bubbles in the electrolyte reservoir.
This does not indicate that the bubbles
caused a problem while the sensor was
actually in use. Remove the
Wash off electrolyte with distilled water.