Reminder:
You save the selected parameter values
by pressing the
STOP
key.
1. Complete the mechanical settings
and adjust the carriage to the patient
as described in section 4.2.
2. Press the
MENU key
on the pro-
gramming unit to access level 1 (M1).
3. Press the
parameter key
for
internal
rotation
or
external
rotation
and press the
+/-
keys
to set the carriage to the
desired rotation position to be
maintained during the isolated
abduction/adduction movements.
4. Next set the desired range of motion
for adduction and abduction by first
pressing the
parameter key
for
adduction
and set the values
with the
+/- keys
.
5. Set the abduction value in the same
way
.
6. Press the
MENU key
on the pro-
gramming unit to access level 3 (M3).
7. Press the Motor B ON/OFF parame-
ter key
to activate the parame-
ter.
8. Deactivate motor B by pressing the
parameter key
Motor B ON/OFF
again or the "–" (minus) key. The
check mark in the circle next to the
symbol must have disappeared.
9. Save the settings with the
STOP key
,
set the carriage to the starting
Note!
- For an exclusive
adduction/abduction protocol, motor
B for the rotation movement must be
turned off as described in steps 6 to
8 below.
- Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
5.6.1 Isolated
adduction/abduction
5.6 Examples of use/
sample protocols
70