General Characteristics
132
Portable Bedside Capnograph/Pulse Oximeter
General Characteristics
Item
Value
Unit Dimensions
167mm(h) x 220mm(w) x 192mm(d)
(6.6in (h) x 8.7in (w) x 7.6in (d))
Unit Weight
3.6kg (7.94lb)
Operating Temperature
0
°
C to 35
°
C (32
°
F to 95
°
F)
Operating Pressure and Altitude
Pressure: 430 mmHg to 795 mmHg
Altitude: -381m to 3000m (-1,250 feet to 9,843 feet)
Operating Humidity
10% to 95% non-condensing
Storage and Transport
Temperature
Until lower limit of -20
°
C (-4°F)
Up to upper limit of 60
°
C (140°F)
Storage and Transport Pressure
and Altitude
Pressure: 430 mmHg to 795 mmHg
Altitude: -381m to 4572m (-1,250 feet to 15,000 feet)
Storage and Transport Humidity
10% -95% non-condensing
Packaged Dimensions
315mm(h) x 340mm(w) x 285mm(d)
(12.4in (h) x 13.4in (w) x 11.2in (d))
Packaged Weight
6.6kg (14.55lb)
Equipment Classification
Item
Value
Types of Protection against Electric
Shock
Class 1
Degree of Protection against
Electric Shock
Type BF defibrillation proof applied parts
Mode of Operation
Continuous
Degree of Protection against
Ingress of liquids
IEC 60601-1, sub-clause 11.6.5 for class IPX1 Drip-proof
equipment*
* IPx1 indicates that the device is protected against harmful effects of vertically falling drops of water.
Compliance
This product is designed to conform to the following standards:
IEC/EN60601-1
UL 60601-1
CSA C22.2 No 601.1-M90
IEC/EN60601-1-2 Class A Radiated and Conducted Emission
IEC 60601-1-8 (Audible and Visual Alarms)
ISO 80601-2-55 (Capnography)
ISO 80601-2-61 (Pulse Oximetry)
IEC 60601-2-49 Particular requirements for the safety of multifunction patient monitoring equipment
Electromagnetic Immunity
The monitor is suitable for use in the specified electromagnetic environment. The user of the monitor should
assure that it is used in an electromagnetic environment as described below.