Outputs
130
Portable Bedside Capnograph/Pulse Oximeter
Pin
Assignment
Pin
Assignment
1
Ground
9
Ground
2
Ch 1 Signal
10
Ground
3
Ground
11
Ch 5 Signal
4
Ch 2 Signal
12
Ground
5
Ground
13
Ch 6 Signal
6
Ch 3 Signal
14
Ground
7
Ground
15
Ch 7 Signal
8
Ch 4 Signal
Nurse Call
Normally Open/ Normally Closed Relay
Rated Carrying current : 2A
Max Allowable Current : 2A
Max Allowable Voltage : 24V DC
Contact Capacity: 2A @ 24V DC.
1/8” stereo phono-jack
Pin out of mating stereo phono plug
N1 - N2: Normally Closed relay
N1 - N3: Normally Open relay