orfit RAYCAST 33221 Instructions For Use Download Page 1

 

 

Page 1 of 2 

 

INSTRUCTIONS FOR USE  

 

 

RAYCAST

®

 HIGH PRECISION BASE 

PLATE-TO-COUCH TOP FIXATION 
DEVICES 

2-PIN BAR FOR SIEMENS 
MAGNETOM FAMILY MRI-
COUCHES (MR SAFE) 

 

Article No. : 

33221

 

A.

 

GENERAL PRODUCT INFORMATION 

This product is a fixation device to attach Raycast High Precision 
base plates to different types of couches of the Siemens 
Magnetom family. . 

B.

 

PRODUCT DESCRIPTION 

This fixation device can be used on couches of different types of 
MRI scanners of the Siemens Magnetom family such as Skyra, 
Spectra and Aera with wide T groove (see below for serial 
number*), Vida and Sola.  
It features two positioning elements that can be inserted in the 
lateral cut-outs on both sides of these Siemens  MRI couches. It 
also features two pins on top that can be used to index 
hardware (base plates, shoulder positioning device, leg indexing 
system, etc.) directly or via an adapter plate to the couch. The 
2-pin bar can be converted into a 3-pin bar by adding the 
additional pin that is delivered with the 2-pin bar. Information 
on these other parts, such as the base plates, can be found in 

the respective ‘instructions for use’.

 

 

*The MAGNETOM systems including couches with the wide T 
groove have a serial number higher than or equal to:  

 

Aera: 42466 

 

Skyra fixed table: 46160 

 

Skyra mobile table 46165 

 

Skyra fit: 70025 

 

Spectra: 72047 

 

C.

 

POSITIONING ON A COUCH 

The two blocks of this fixation device fit directly in the cut-outs 
of the Siemens Magnetom MRI couches mentioned above.  

 

 

To position this 2-pin bar on such a couch, gently push the bar 
into the couch cut-outs. The bar itself should be flat on the 
surface of the couch. The 2-pin bar will be able to slide in a 
longitudinal direction. Always make sure that the bar is well 
positioned on the couch before indexing other hardware to the 
2-pin bar. 

This 2-pin bar cannot be used to position accessories in 
overhang.  

To remove the bar, hold the two ends of the bar and pull 
upwards until the bar snaps off the couch. Never pull in the 
middle of the bar as this will push the positioning blocks even 
harder onto the couch. 

D.

 

STORAGE 

Always store the product in a safe place to prevent it from 
getting damaged. 
Store the product b10°C (50°F) and 40°C (122°F). 
 

E.

 

PROPERTIES 

E.1 Physical Properties 

Dimensions : L 545 mm x W 24 mm x H 23.5 mm 
Weight: 0.1 kg 
 

E.2 Material Properties 
 

This product is made entirely of electrically non-conductive, 
non-metallic and non-magnetic materials and is MR safe. 
 
 

 

 
 
Warning: keep metal wrenches which might be supplied in the 
package out of the MRI room.  
 

E.3 Dosimetric Properties 

Attenuation factor (± 0.15 %) 

6 MV 

2.67 % 

15 MV 

2.00 % 

H

2

O equivalence  

7.72 mm 

Note:  Use  these  numbers  as  a  guidance  only.  Perform  the 
measurements again in your department to verify these results.  

 

 

 

 

 

Reviews: