DP-100SL
13
Site Preparation Guide
Chapter 3
Facility Physical Requirements
Safety Requirements
The DP-100SL should be installed in a room that can be isolated to
safeguard against laser hazards. These hazards can only arise during
periodic maintenance and service, in specific situations (hood open and
interlock neutralized). The safeguards described below are required during
such service procedures, and do not have to be activated during normal
usage of the system.
Preventing Access to the Imager Room during Service
The imager room door should have a lock that can be locked from the
inside to prevent access to the room. A box with a key to be used for entry
in case of emergency should be located next to the door, on the outside.
Note
The customer is responsible for ensuring conformance with local safety
regulations. As such, the requirements below can only be considered to be
recommendations.
!
Warning
When the DP-100SL cover doors are removed (servicing mode), the laser
becomes a Class IV hazard. In this situation, the laser beams diffused
reflection is extremely hazardous. Goggles must be worn. Service must
only be performed by an Orbotech qualified engineer.
Summary of Contents for DP-100SL
Page 1: ...DP 100SL Site Preparation Guide October 2003 Revision A ...
Page 2: ......
Page 14: ......
Page 18: ......
Page 25: ...DP 100SL 19 Site Preparation Guide Figure 3 2 System Dimensions Front View ...
Page 26: ...20 Chapter 3 Facility Physical Requirements Figure 3 3 System Dimensions Side View ...
Page 27: ...DP 100SL 21 Site Preparation Guide Figure 3 4 Site Dimensions Standard Configuration ...
Page 32: ......
Page 40: ......
Page 48: ......