20
|
INSTRUCTIONS FOR USE
OM 402LC
Items for “Limits” and “Analog output” are acces-
sible only if incorporated in the instrument.
!
Display value for the beginning of the AO range > MIN A.O. = 0
Example
TYP. A.O.
0
Assigning the display
value to the beginning of
the AO range
MIN A.O.
Assigning the display
value to the beginning of
the AO range
- range of the setting is -99999…999999
DEF
= 0
0
MAX A.O.
MIN A.O.
Displa
y
ed only with op
tions >
Analog output
TYP. A.O.
Setting the type of analog
output
Menu
Range
Description
0-20mA
0…20 mA
Er.4- T
4…20 mA
signaling interrupted current loop and displaying
an error message (<3,6 mA)
4-20 T
4…20 mA
signaling broken current loop
E. 4-20mA
4…20 mA
with indication of error statement (<3,6 mA)
4-20mA
4…20 mA
0-5mA
0…5 mA
0-2 V
0…2 V
0-5 V
0…5 V
0-10 V
0…10 V
+-10 V
±10 V
DEF
= 4…20 mA
Type of analog output - 0...10 V > TYP. A.O. = 0-10 V
Example
4-20mA
0-2 V
0-5mA
0-10 V
0-5 V
MIN A.O.
5
light
light
SETTING
0-20mA
Er.4- T
4-20 T
Er.4-20
0-10 V
+-10 V
• • •