background image

 

Operator

’s Manual

 

    

optris

®

 PI 

 
 
 

 

400i/ 450i/ 450i G7/ 640/ 640 G7/ 05M/ 08M/ 1M

 

 

Infrared camera

 

Summary of Contents for PI 640

Page 1: ...Operator s Manual optris PI 400i 450i 450i G7 640 640 G7 05M 08M 1M Infrared camera ...

Page 2: ...Optris GmbH Ferdinand Buisson Str 14 13127 Berlin Germany Tel 49 30 500 197 0 Fax 49 30 500 197 10 E mail info optris global Internet www optris global ...

Page 3: ...s Table of contents 3 1 General Notes 7 1 1 Intended use 7 1 2 Warranty 8 1 3 Scope of delivery 9 1 4 Maintenance 9 1 4 1 Cleaning 9 1 5 Model overview 10 2 Technical Data 11 2 1 General specifications 11 2 2 Electrical specifications 15 ...

Page 4: ... Changing the lens 35 3 3 Fixing the focus of the lens only for PI 05M 08M 1M 36 3 4 Mounting accessories optional 38 3 5 High temperature accessories 39 3 5 1 CoolingJacket 39 3 5 2 CoolingJacket Advanced 41 3 5 3 Laminar air purge for CoolingJacket 45 3 5 4 Outdoor protective housing 46 4 Electrical Installation 47 ...

Page 5: ... 2 Example for a Fail Safe monitoring of the PI with a PLC 55 4 3 USB cable extension 57 5 IRmobile App 59 6 Software PIX Connect 61 6 1 Installation and initial start up 62 6 2 Software window 64 6 2 1 Basis features of the software PIX Connect 66 7 Basics of Infrared Thermometry 69 8 Emissivity 75 8 1 Definition 75 ...

Page 6: ...le of emissivity for metals 80 Appendix B Table of emissivity for non metals 82 Appendix C Quick start for serial communication 83 Appendix D Interprocess Communication IPC 85 Appendix E PIX Connect Resource Translator 86 Appendix F Wiring diagrams PIF 87 Appendix G Declaration of Conformity 91 ...

Page 7: ...ser to do a comfortable detailed analysis with the software PIX Connect The PI is a precise instrument and contains an extremely sensitive infrared detector and a high quality lens The alignment of the camera to intensive energy sources e g devices which emit laser radiation or reflections of such equipment can cause an irreparable defect of the infrared detector This is also valid if the camera i...

Page 8: ...e After the warranty is expired the manufacturer guarantees additional 6 months warranty for all repaired or substituted product components Warranty does not apply to damages which result from misuse or neglect The warranty also expires if you open the product The manufacturer is not liable for consequential damage or in case of a non intended use of the product If a failure occurs during the warr...

Page 9: ...ss interface cable incl terminal block 1 m Software package PIX Connect Operators manual Aluminum case PI 450i 450i G7 640 640 G7 only robust hard transport case IP67 1 4 Maintenance Never use cleaning compounds which contain solvents neither for the lens nor for the housing 1 4 1 Cleaning Blow off loose particles using clean compressed air The lens surface can be cleaned with a soft humid tissue ...

Page 10: ...Scanning mode PI 640 IR 20 to 900 C 200 to 1500 C optional 8 14 µm 32 Hz Pin sharp radiometric recordings in real time PI 640 G7 IR 200 to 1500 C 150 to 900 C 7 9 µm 32 Hz Measurement of glass with Line Scanning mode PI 05M IR 900 to 2450 C 500 540 nm Up to 1 kHz Measurement of metallic surfaces graphite or ceramics with short wavelengths PI 08M IR 575 1900 C 780 820 nm Bis 1 kHz Measurement of me...

Page 11: ... Aluminum anodized Dimensions PI 400i PI 450i 450i G7 46 x 56 x 68 77 mm depending on lens and focus position PI 640 640 G7 46 x 56 x 76 100 mm depending on lens and focus position PI 640 microscope optics 46 x 56 x 119 126 mm depending on focus position PI 05M PI 08M PI 1M 46 x 56 x 88 129 mm depending on lens and focus position Weight PI 400i PI 450i 450i G7 237 251 g depending on lens PI 640 64...

Page 12: ...standard 3 m 5 m 10 m 20 m Vibration1 IEC 60068 2 6 sinus shaped IEC 60068 2 64 broadband noise Shock1 IEC 60068 2 27 25 G and 50 G 1 Used standards for vibration and shock Figure 1 Used standards Stress program camera in operation ...

Page 13: ...ation 600 Shocks 100 Shocks each direction Shock half sinus 50 G testing Ea 50 G acc IEC 60068 2 27 Acceleration 490 m s2 50 G Pulse duration 11 ms Number of directions 6 3 axes with two directions each Duration 18 Shocks 3 Shocks each direction Vibration sinus shaped testing Fc acc IEC60068 2 6 Frequency range 10 500 Hz Acceleration 29 42 m s2 3 G Frequency change 1 Octave min ...

Page 14: ...C60068 2 64 Frequency range 10 2000 Hz Acceleration 39 3 m s2 4 01 GRMS Frequency spectrum 10 106 Hz 0 9610 m s2 2 Hz 0 010 G2 Hz 106 150 Hz 6 dB Octave 150 500 Hz 1 9230 m s2 2 Hz 0 020 G2 Hz 500 2000 Hz 6 dB Octave 2000 Hz 0 1245 m s2 2 Hz 0 00126 G2 Hz Number of axes 3 Duration 3 h 3 x 1 h ...

Page 15: ... PIF AI Input Standard Process Interface PIF in 0 10 V Emissivity ambient temperature reference temperature uncommitted value flag control triggered snapshots triggered recording triggered linescanner triggered event grabber reset peak value hold switch temperature range Appendix F Wiring diagrams PIF DI Digital Input Standard Process Interface Flag control triggered snapshots triggered recording ...

Page 16: ...accuracy 3 2 C or 2 Thermal sensitivity NETD 75 mK4 with 29 53 and 80 0 1 K4 with 18 40 mK4 with 29 53 and 80 60 mK4 with 18 150 mK Tobj 650 C with 29 53 80 175 mK Tobj 650 C with 18 Warm up time 10 min Emissivity 0 100 1 100 Software PIX Connect 1 Accuracy statement effective from 150 C 2 The sighting range is used to align the G7 cameras at ε 1 and activation of the extended temperature range a ...

Page 17: ... 64 F 0 8 Microscope lens FOV 12 x 9 F 1 1 System accuracy 3 2 C or 2 Thermal sensitivity NETD 75 mK with 33 60 and 90 85 mK with 15 130 mK Tobj 650 C with 33 60 90 150 mK Tobj 650 C with 15 Warm up time 10 min Emissivity 0 100 1 100 Software PIX Connect 1 Accuracy statement effective from 150 C 2 The sighting range is used to align the G7 cameras at ε 1 and activation of the extended temperature ...

Page 18: ... 4 FOV 382x288 px 13 x 10 FOV 764x480 px 26 x 16 F 1 4 39 x 25 F 1 4 FOV 382x288 px 13 x 10 20 x 15 FOV 764x480 px 9 x 5 F 2 8 13 x 8 F 2 4 26 x16 F 1 4 39 x 25 F 1 4 FOV 382x288 px 4 x 3 7 x 5 13 x 10 20 x 15 System accuracy 3 For object temperature 2000 C 1 of reading for 27 32 80 Hz 1 5 of reading for 1 kHz For object temperature 2000 C 2 of reading for 27 32 80 Hz 2 5 of reading for 1 kHz For ...

Page 19: ... 1 75 C start temperature for optics with focal length f 50 mm f 75 mm 2 An additionally purchased lens for the PI 05M 08M 1M camera comes with the corresponding protective tube 3 At an ambient temperature of 25 C 4 Specified NETD value applies to all frequencies ...

Page 20: ...ics Figure 2 The turning out of the optics leads to the focus setting near and the turning in of the lens to the focus setting infinity Figure 2 Focusing by turning the exterior lens ring of camera Figure 3 PI 05M PI 08M PI 1M For the PI 05M PI 08M and PI 1M camera you must first unscrew the protective tube in order to be able to focus the camera Figure 3 ...

Page 21: ...es We offer lenses for close standard distances and large distances Different parameters are important if using infrared cameras They display the connection between the distance of the measured object and the size of the pixel Table 2 Figure 4 Measurement field of the infrared camera optris PI representing the 23 x 17 lens ...

Page 22: ...amples showing what spot sizes and pixel sizes will be reached in which distance For individual configuration there are different lenses available Wide angle lenses have a radial distortion due to their large opening angle the software PIX Connect has an algorithm which corrects this distortion As an alternative to the tables below the optics calculator can also be used on the optris website https...

Page 23: ...Technical Data 23 Table 2 Note The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance ...

Page 24: ... 24 Note The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance ...

Page 25: ...Technical Data 25 Note The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance ...

Page 26: ...he accuracy of measurement can be outside of the specifications for distances below the defined minimum distance 1 PI 05M is only available with OF25 optics and PI 08M is only available with OF16 and OF25 optics ...

Page 27: ... Note The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance 1 PI 05M is only available with OF25 optics and PI 08M is only available with OF16 and OF25 optics ...

Page 28: ...tric M4 thread holes on the bottom side 6 mm depth and can be installed either directly via these threads or with help of the tripod mount also on bottom side The tightening torque of the M4 screws for mounting the PI camera should be between 1 1 5 Nm and must not exceed 2 Nm ...

Page 29: ...Mechanical Installation 29 Figure 5 PI 400i PI 450i PI 450i G7 dimensions mm ...

Page 30: ... 30 Figure 6 PI 640 PI 640 G7 optics 29 33 53 60 dimensions mm ...

Page 31: ...Mechanical Installation 31 Figure 7 PI 640 PI 640 G7 optics 13 15 dimensions mm ...

Page 32: ... 32 Figure 8 PI 640 PI 640 G7 optics 80 90 dimensions mm ...

Page 33: ...Mechanical Installation 33 Figure 9 PI 640 microscope optics 10 12 dimensions mm ...

Page 34: ... 34 Figure 10 PI 05M PI 08M PI 1M dimensions mm ...

Page 35: ...st be turned off see Figure 3 Figure 11 Change lens for PI 160 2xx 4xxi 640 Figure 12 Change lens for PI 05M 08M 1M Figure 13 Inserting the lens To get the best possible measurements when inserting the lens into the camera body make sure that the label on the lens is screwed in at the same height as the label from the housing see Figure 13 1 An additionally purchased lens for the PI 05M 08M 1M cam...

Page 36: ...s unscrew the protective tube of the camera see Figure 3 There are three small holes on the lens Take the three screws that are included and attach them to the three holes The focus of the lens is now fixed Alternatively the two knurled screws supplied can also be used Figure 14 Lens for PI 05M 08M 1M Figure 15 Fixing the focus for PI 05M 08M 1M ...

Page 37: ...Mechanical Installation 37 Figure 16 Focusing screws for focus ring ...

Page 38: ... 38 3 4 Mounting accessories optional Figure 17 Mounting base stainless steel adjustable in 2 axes Part No ACPIMB Figure 18 Protective housing stainless steel Incl Mounting base Part No ACPIPH ...

Page 39: ...ure accessories 3 5 1 CoolingJacket The IR camera can be used at ambient temperature up to 50 C up to 70 C with PI 450i PI 450i G7 For higher temperatures up to 180 C the CoolingJacket is provided For detailed information see installation manual ...

Page 40: ... 40 Figure 19 CoolingJacket Dimensions ...

Page 41: ...5 2 CoolingJacket Advanced The CoolingJacket Advanced is available as Standard Version and Extended Version The IR camera can be used at ambient temperature up to 50 C up to 70 C with PI 450i PI 450i G7 For higher temperatures up to 315 C the CoolingJacket Advanced is provided For detailed information see installation manual ...

Page 42: ... 42 Standard Version Figure 22 CoolingJacket Advanced Part No ACPIxxxCJAS Standard Version Dimensions ...

Page 43: ...strial PIF or the USB Server Gigabit and industrial PIF Both PI Netbox and industrial PIF or USB Server Gigabit and industrial PIF can be integrated in the CoolingJacket Figure 23 Cooling Jacket Advanced Extended Version with PI Netbox and industrial PIF Figure 24 Cooling Jacket Advanced Extended Version with USB Server and industrial PIF ...

Page 44: ... 44 Figure 25 CoolingJacket Advanced Part No ACPIxxxCJAE Extended Version Dimensions ...

Page 45: ...e available One for standard IR camera applications Part No ACCJAAPLS and the other for line scanning applications Part No ACCJAAPLL Those two versions are fitting to all focusing units with production date 01 2018 A protective window 67 x 3 mm has to be ordered separately If you like to mount the air purge on an older CJ the focusing unit should be exchanged to the current version ...

Page 46: ...ctive housing The outdoor protective housing can be used for any PI camera lenses up to 90 FOV In addition the industrial PIF can be installed as an accessory without housing Also available for CSlaser LT or CTlaser LT For detailed information see installation manual Figure 26 Outdoor protective housing for PI camera USB server and industrial PIF ...

Page 47: ...n At the back side of the PI there are the two connector plugs The left plug is for the USB cable The right connector plug is only used for the process interface Figure 27 Backside of the camera with connectors 1 Plug for USB cable 2 Plug for PIF cable ...

Page 48: ...process The signal level is always 0 10 V DI 24 V The process interface can be activated choosing the following options Analog Input AI Emissivity ambient temperature reference temperature uncommitted value flag control triggered recording triggered snapshots triggered linescanner triggered event grabber reset peak value hold switch temperature range Analog Output AO Main measure area measure area...

Page 49: ... inputs and outputs Name Description max range1 status AI Analog input 0 10 V 2 DI Digital input active low 0 0 6 V 24 V AO Analog output Alarm output 0 10 V 0 10 V 1 Depending on supply voltage for 0 10 V on the AO the PIF has to be powered with min 12 V 2 The AI is designed for max 24 V the voltage level above 10 V is not interpreted ...

Page 50: ... 50 4 1 1 PIN allocation Figure 29 Rear side of the camera USB PIF 1 VCC 1 INT 2 GND 2 SDA I C 3 SCL I C 4 D 4 DGND 5 D 5 3 3 V Out ...

Page 51: ...oprietary PIF cable in the menu Tools Configuration Device PIF in the PIX Connect software is necessary Figure 30 Support proprietary PIF cable Consider that the input of the PIF is not protected if there is a direct PIF connection A voltage 3 V on the INT pin will destroy the device 1 We recommend using only a switching contact between INT and DGND as external hardware button relay ...

Page 52: ...cess is available connection box with IP65 5 m 10 m or 20 m standard or high temperature cable for camera connection terminal for process integration Appendix F Wiring diagrams PIF Pin assignment PIF cable industrial process interface Figure 31 Connections of the industrial Process Interface GREY Interrupt GREEN SCL I C YELLOW SDA I C WHITE 3 3 V BROWN GND SHIELD GND ...

Page 53: ...pen closed green LED on 0 30 V 400 mA 1 depending on supply voltage for 0 20 mA on the AO the PIF has to be powered with min 5V 1 5 working resistance 0 021 24 V Example RLoad 500 ohm Umin 1 5 500 0 021 12 V RLoad 100 ohm Umin 1 5 100 0 021 3 6 V min 5 V 2 the AI is designed for max 24 V the voltage level above 10 V is not interpreted 3 active if AO1 2 or 3 is are programmed as alarm output The al...

Page 54: ...onstant of the fail safe is 1 5 seconds Controlled conditions on camera and software Standard Process interface ACPIPIF Industrial Process interface ACPIPIFMACBxx Interruption USB cable to camera Interruption data cable camera PIF Interruption power supply PIF Shut down of PIX Connect software Crash of PIX Connect software Fail Safe Output 0 V at analog output AO open contact fail safe relay green...

Page 55: ...a PLC Figure 32 Fail Safe monitoring states Fail Safe monitoring states 1 Breakdown of PIF power supply 4 Malfunction of PI 2 Cable break of fail safe cable 5 Breakdown of PI power supply Interruption of USB cable 3 Interruption of cable PI PIF 6 Malfunction of PIX Connect software ...

Page 56: ...oring states 1 Breakdown of PIF power supply 5 Malfunction of PI 2 Cable break of fail safe cable 6 Breakdown of PI power supply Interruption of USB cable 3 Short circuit of fail safe cable 7 Malfunction of PIX Connect software 4 Interruption of cable PI PIF ...

Page 57: ... cable length is 20 m For greater distances between PI and computer or for stand alone solutions the optional PI NetBox or the USB Server Gigabit is provided Figure 34 Ethernet direct communication with PI Netbox Figure 35 Ethernet network communication with PI Netbox ...

Page 58: ... 58 Figure 36 Stand Alone operation with PI Netbox Figure 37 USB Server Gigabit ...

Page 59: ...nection to the device Part No ACPIIACM micro USB or ACPIIACC USB C With IRmobile you can monitor and analyze your infrared temperature measurement directly on a connected smartphone or tablet All you need is an Optris infrared camera This app runs on most Android 5 or higher devices with a micro USB port or USB C port that supports USB OTG On The Go The app is easy to use After you have connected ...

Page 60: ... with automatic hot and coldspot search Changing the color palette scaling and temperature range Change of temperature unit Celsius or Fahrenheit Setting of temperature range scaling Manual Min Max 3 sigma Creating a snapshot Integrated simulator IRmobile supported for Optris IR cameras PI and Xi series Optris pyrometers Compact series high performance series and video thermometers For Android 5 o...

Page 61: ...sc with at least 30 MByte of free space At least 128 MByte RAM A detailed description is provided in the software manual on the USB stick See also Help menu in the PIX Connect software Help Documentation Alternatively the software can also be downloaded via the Optris website under the following link https www optris global pix ...

Page 62: ... finished The installation wizard places a launch icon on the desktop and in the start menu Start Programs Optris GmbH PIX Connect 3 To connect the camera to the PC plug the USB cable to the camera first Afterwards connect it with the PC to disconnect the camera and the computer remove the USB cable from the computer first and then disconnect it from the camera 4 Start the software At the initial ...

Page 63: ...r After the calibration files have been installed the live image from the camera is shown inside a window on your PC screen 6 Choose the desired language in the menu Tools Language 7 Adjust the focus of the image by turning the exterior lens ring at the camera ...

Page 64: ... 64 6 2 Software window Figure 39 Software window 1 2 2 2 3 4 5 6 7 8 9 10 11 ...

Page 65: ...rm settings Bar showing the defined temperature thresholds for low alarm value blue arrow and high alarm value red arrow The color of numbers within control displays changes to red when temp above the high alarm value and to blue when temp below the low alarm value 6 Temperature time diagram Shows the temperature curves over time for selectable region of interest ROI 7 Histogram Shows the statisti...

Page 66: ...n different windows Compatible with Windows 7 8 and 10 High level of individualization for customer specific display Various language option including a translation tool Temperature display in C or F Different layout options for an individual setup arrangement of windows toolbar Range of individual measurement parameter fitting for each application Adaption of thermal image mirror rotate Individua...

Page 67: ...Extensive online and offline data analysis Analysis supported by measurement fields hot and cold spot searching image subtraction Real time temperature information within main window as digital or graphic display line profile temperature time diagram Slow motion repeat of radiometric files and analysis without camera being connected Editing of sequences such as cutting and saving of individual ima...

Page 68: ...M Ports and DLL Adjustment of thermal image via reference values Temperature data analysis and documentation Triggered data collection Radiometric video sequences ravi radiometric snapshots tiff Text files including temp information for analysis in Excel csv dat Data with color information for standard programs such as Photoshop or Windows Media Player wmv tiff Data transfer in real time to other ...

Page 69: ...iation Searching for new optical material William Herschel by chance found the infrared radiation in 1800 Figure 40 William Herschel 1738 1822 He blackened the peak of a sensitive mercury thermometer This thermometer a glass prism that led sun rays onto a table made his measuring arrangement With this he tested the heating of different colors of the spectrum Slowly moving the peak of the blackened...

Page 70: ...e length ranging between 1 µm and 20 µm The intensity of the emitted radiation depends on the material This material contingent constant is described with the help of the emissivity which is a known value for most materials 8 Emissivity Infrared thermometers are optoelectronic sensors They calculate the surface temperature on the basis of the emitted infrared radiation from an object The most impo...

Page 71: ...nearization signal processing The specifications of the lens decisively determine the optical path of the infrared thermometer which is characterized by the ratio Distance to Spot size The spectral filter selects the wavelength range which is relevant for the temperature measurement The detector in cooperation with the processing electronics transforms the emitted infrared radiation into electrica...

Page 72: ...rature measurements of moving or overheated objects and of objects in hazardous surroundings very fast response and exposure times measurement without inter reaction no influence on the measuring object non destructive measurement long lasting measurement no mechanical wear ...

Page 73: ...Basics of Infrared Thermometry 73 Figure 43 Non contact thermometry ...

Page 74: ...ring of electronic cabinets R D of electronics R D of electronic parts Process control extruding plastic parts Process control manufacturing solar modules Process control at calendering R D of mechanical parts Monitoring of cables ...

Page 75: ...ial of the measuring object The emissivity ε Epsilon is used as a material constant factor to describe the ability of the body to emit infrared energy It can range between 0 and 100 A blackbody is the ideal radiation source with an emissivity of 1 0 whereas a mirror shows an emissivity of 0 1 Figure 44 Composition of IR radiation I IR radiation ε Emission ρ Reflection τ Transmission ε ρ τ 1 ...

Page 76: ...eal temperature assuming the measuring object is warmer than its surroundings A low emissivity reflective surfaces carries the risk of inaccurate measuring results by interfering infrared radiation emitted by background objects flames heating systems chamottes To minimize measuring errors in such cases the handling should be performed very carefully and the unit should be protected against reflect...

Page 77: ...esponds to the actual temperature If you monitor temperatures of up to 380 C you may place a special plastic sticker emissivity dots Part No ACLSED onto the measuring object which covers it completely Figure 46 Plastic sticker at metal surface Set the emissivity to 0 95 and take the temperature of the sticker Afterwards determine the temperature of the adjacent area on the measuring object and adj...

Page 78: ...e the temperature of the colored surface Afterwards determine the temperature of a directly adjacent area and modify the emissivity until the measured value corresponds to the temperature of the colored surface Figure 47 Shiny metal surface left and blackened metal surface right CAUTION On all three methods the object temperature must be different from ambient temperature ...

Page 79: ...erage values only The actual emissivity of a material depends on the following factors temperature measuring angle geometry of the surface thickness of the material constitution of the surface polished oxidized rough sandblast spectral range of the measurement transmissivity e g with thin films Figure 48 Adjustment of the emissivity in the software PIX Connect menu Tools Configuration Device ...

Page 80: ... 80 Appendix A Table of emissivity for metals ...

Page 81: ...Appendix A Table of emissivity for metals 81 ...

Page 82: ... 82 Appendix B Table of emissivity for non metals ...

Page 83: ...communicate via a serial COM Port interface This can be a physical COM Port or a virtual COM Port VCP It must be available on the computer where the PIX Connect software is installed Setup of the interface 1 Open the Configurations dialog and enter the tab External Communication to enable the software for the serial communication 2 Select the mode COM Port and choose the appropriate port ...

Page 84: ...t 8N1 These parameters are used in many other communication devices too The other station must support 8 bit data 4 Connect the computer with the communication device If this is a computer too use a null modem cable Command list The command list is provided on the USB stick and in the PIX Connect software Help SDK Every command must expire with CR LF 0x0D 0x0A ...

Page 85: ...nect software Imager exe only A dynamic link library ImagerIPC2 dll provides the interprocess communication IPC for other attached processes The DLL can be dynamically linked into the secondary application Or it can be done static by a lib file too Both Imager exe and ImagerIPC2 dll are designed for Windows 7 8 10 only The application must support call back functions and polling mode The ImagerIPC...

Page 86: ...cumentation on http msdn microsoft com en us goglobal bb688096 aspx If desired the localization process can be illustrated in detail Also the resizing of buttons or other visible resources and the support of right to left languages are supported Experts who have the appropriate tools should handle it Nevertheless we have developed the small tool Resource Translator to make the translation of the r...

Page 87: ...agrams PIF 87 Appendix F Wiring diagrams PIF Analog Output The maximum load impedance is 500 Ohm The analog output can be used as a digital output too The current value for no alarm and alarm on is set within the software ...

Page 88: ...igital Input Figure 49 Digital input The digital input can be activated with a button to the PI GND Pin or with a low level CMOS TTL signal Low level 0 0 6 V High level 2 24 V Example Button Figure 50 Button ...

Page 89: ... voltage range 0 10 V Figure 51 Analog input Relay output at industrial PIF Part No ACPIPIFMACBxx The analog output must be set to Alarm The range for AO1 AO3 can be set in the software no alarm 0 4 mA alarm 10 20 mA REL1 3 DO1 DO3 Umax 30 VDC Imax 400 mA ...

Page 90: ... 90 Figure 52 Relay output at industrial PIF ...

Page 91: ...Appendix G Declaration of Conformity 91 Appendix G Declaration of Conformity ...

Page 92: ...optris PI MA E2020 05 A ...

Reviews: