Technical Data
11
-
2.4 Air Purge Laminar
The lens must be always kept clean from dust, smoke, fumes, and other contaminants in order to avoid reading
errors. These effects can be reduced by using an air purge laminar. Make sure to use oil-free, technically clean
air only.
The needed amount of air (approx. 20...100 l/min.) depends on the application and the installation
conditions on-site. Pipe connection of air purge laminar is 6x4 mm.
2.5 Protective window
Protective window
Description
Spectral range
Transmission
1)
Germanium
Protective window (50,8 x 3 mm/ Ge) for PI 450i G7, 640i G7
7,9 µm
0.96
Protective window (50,8 x 3 mm/ Ge) for PI 400i, 450i, 640i
8-14 µm
0.92
Zinc sulfide
Protective window (50,8 x 3 mm/ ZnS) for PI 450i G7, 640i G7
7,9 µm
0.93
Protective window (50,8 x 3 mm/ ZnS) for PI 400i, 450i, 640i
8-14 µm
0.91
Protective window (50,8 x 3 mm/ ZnS) for LT models of CTlaser-, CSlaser serie
8-14 µm
0,91
Borofloat
Protective window (50,8 x 3 mm/ Borofloat 33) for PI 1M, 08M, 05M
0,85-1,1 µm, 780-820 nm,
500-540 nm
0.92
1)
The displayed values are standard values and can vary between different delivery units.
When using the foil
[part number: ACOPHSF]
, the transmission must be set on 0.85 (deviations possible).
Summary of Contents for Outdoor protective housing
Page 8: ... 8 2 2 Dimensions Figure 1 Dimensions mm with wall mount ...
Page 9: ...Technical Data 9 Figure 2 Dimensions for wall mount ...
Page 26: ... 26 Figure 24 Remove air purge laminar ...
Page 32: ... 32 Appendix A Declaration of Conformity ...
Page 33: ......
Page 34: ...optris Outdoor protective housing MA E2022 10 A ...