4.1 Process interface
The process interface (electronics within cable as well as industrial interface) must be powered
separately (5-24 VDC). Before switching on the power the PIF cable must be connected to the
camera.
The PI is equipped with a process interface (cable with integrated
electronics and terminal block), which can be programmed via the
software as an Analog Input (AI) and Digital Input (DI)
in order to
control the camera or as an Analog Output (AO) in order to control
the process. The signal level is always 0-10 V (DI = 24 V).
The process interface can be activated choosing the following options:
Analog Input (AI):
Emissivity, ambient temperature, reference temperature, uncommitted value, flag control, triggered recording,
triggered snapshots, triggered linescanner, triggered event grabber, reset peak-/value-hold, switch temperature
range
Analog Output (AO):
Main measure area, measure area, internal temperature, flag status, recording status, line scan status, alarm,
frame sync, fail-safe, external communication
Digital Input (DI):
Flag control, triggered snapshots, triggered recording, triggered linescanner, triggered event grabber, reset peak-
/value-hold, switch temperature range
Summary of Contents for optris PI 640 G7
Page 32: ...32 Figure 7 PI 160 PI 400i PI 450i PI 450i G7 dimensions mm...
Page 33: ...Mechanical Installation 33 Figure 8 PI 200 PI 230 dimensions mm...
Page 34: ...34 Figure 9 PI 640 PI 640 G7 optics 29 33 53 60 dimensions mm...
Page 35: ...Mechanical Installation 35 Figure 10 PI 640 PI 640 G7 optics 13 15 dimensions mm...
Page 36: ...36 Figure 11 PI 640 PI 640 G7 optics 80 90 dimensions mm...
Page 37: ...Mechanical Installation 37 Figure 12 PI 640 microscope optics 10 12 dimensions mm...
Page 38: ...38 Figure 13 PI 05M PI 08M PI 1M dimensions mm...
Page 41: ...Mechanical Installation 41 Figure 19 Focusing screws for focus ring...
Page 44: ...44 Figure 22 CoolingJacket Dimensions...
Page 48: ...48 Figure 28 CoolingJacket Advanced Part No ACPIxxxCJAE Extended Version Dimensions...
Page 62: ...62 Figure 39 Stand Alone operation with PI Netbox Figure 40 USB Server Gigabit...
Page 68: ...68 6 2 Software window Figure 42 Software window 1 2 2 2 3 4 5 6 7 8 9 10 11...
Page 77: ...Basics of Infrared Thermometry 77 Figure 46 Non contact thermometry...
Page 84: ...84 Appendix A Table of emissivity for metals...
Page 85: ...Appendix A Table of emissivity for metals 85...
Page 86: ...86 Appendix B Table of emissivity for non metals...
Page 94: ...94 Figure 55 Relay output at industrial PIF...
Page 95: ...Appendix G Declaration of Conformity 95 Appendix G Declaration of Conformity...
Page 96: ...optris PI MA E2019 06 A...