ACCTRS485USBK / ACCTRS485B                                    RS485-Interface-Kit 

 

 

   

 

 

 

                                                                    ACCTRS485-MAD-ED2016-10-A 

I

 

 

Quick Reference/ Kurzanleitung 

 

Scope of Supply/ Lieferumfang 

 

ACCTRS485USBK 

  RS485/ RS422 

– USB-Adapter (incl. 1m cable and terminal block/                

inkl. 1m Kabel und Anschlussklemmblock) 

  USB extension cable/ USB-Verlängerung (60cm) 

  Software-CD 

  Quick reference/ Kurzanleitung 

 

ACCTRS485B 

  RS485-Interface board/ RS485-Interface-Platine 

  Cable gland/ Kabelverschraubung M12x1,5 

 

 

 

      Installation of the RS485 interface board/ 

  Quick reference/ Kurzanleitung 

 

 

 

 

      Einbau der RS485-Interface-Platine 

  Mounting screws, cable tie/ Schrauben, Kabelbinder 

 

Installation 

 

  Please plug the RS485-interface board into the place provided, which is located beside the display. In the correct position the holes of the 

RS485-interface match with the thread holes of the CT box.  

  Now press the RS485-interface down to connect it with the CT. Use both M3x5 screws for fixing the board.  

  The RS485-interface provides a 2-wire half-duplex mode. Please connect terminal A of the interface with terminal A of the next RS485-

interface and so on. With the B terminals proceed as well. Exchange the blind screw on each CT box by the cable gland. 

  Make sure, that you always connect A to A and B to B, not reverse. You may run up to 32 CT units on one RS485-USB-adapter.              

The 120R-switch has to be turned to ON at one of the connected CT units only. 

  Each CT unit connected to the RS485 needs a different multidrop address (1...32). Please adjust the address by pressing the mode 

button until M xx appears in the display. Using the Up- and Down-keys you can change the shown address (xx). The address can also be 
changed with the software. The setting for baud rate in the software must be the same as on the CT unit. 

 

  Stecken Sie die RS485-Interface-Platine in die dafür vorgesehene Aufnahme im CT, welche sich neben der Anzeige befindet. Die richtige 

Lage ist erreicht, wenn die Schraubenlöcher des RS485-Interface mit denen des CT übereinstimmen. 