Scope of Supply/ Lieferumfang
RS232 interface board incl. terminal block/ RS232 Interface-Platine inkl.
Anschlussklemmblock
Software CD
RS232 interface cable/ RS232 Schnittstellenkabel (1,5 m)
Cable gland/ Kabelverschraubung M12x1,5
Mounting screws, cable tie/ Schrauben, Kabelbinder
Quick reference/ Kurzanleitung
Installation of the RS232 interface board/
Einbau der RS232-Interface-Platine
Installation
Please plug the RS232 interface into the place provided, which is located beside the display. In the correct position the holes of the
interface match with the thread holes of the CT box.
Now press the PCB down to connect it with the CT. Use both M3x5 screws for fixing the board. Exchange the blind screw on the CT box
by the cable gland and install the RS232 interface cable.
Make sure the wiring is correct according to the drawing and designation printed on the interface board.
The CT always needs an external power supply for operation.
Please install the software as described in the operators manual. Follow the software instruction manual. After you have connected the
RS232 cable to your PC and started the software the communication will be established.
NOTE:
Please make sure that the option
Scan non-USB devices
is activated in the software
[Menu: Preferences/ Options].