Chapter 3 - How to...
4. Select the first stereo image. The cross hairs will move to the left. Align the patient to the first stereo
alignment target. Make sure the patient can see the target, that the target is green and that the cross
hairs are close to the center of the pupil.
5. Capture the image either manually or using auto-capture.
l
Auto-capture: Auto-capture is enabled when
Automatic
is shown in green on the touch screen.
The system captures an image when the scan head is correctly aligned. You can press the
screen to manually capture the image. Auto-capture is turned off after capturing an image.
l
Manual capture: Align the scan head into the correct position. Press the screen to capture the
image when the pupil is in the center of the cross hairs and the patient alignment target is
green. You can also capture using the hand control.
6. Select the second stereo image. The cross hairs will move to the right. Align the patient to the second
stereo alignment target. Make sure the patient can see the target, that the target is green and that the
cross hairs are close to the center of the pupil.
7. Capture the image.
8. Check the quality of the images, see
What to do after the image is captured
3.2.7 How to capture SLO angiography images
The device supports the capture of angiography images.
CAUTION
When angiography images are being captured additional space will be required to allow
the injection of the dye and for other individuals involved in the procedure. These
individuals could include medical staff or patient helper.
Do not inject the patient until the appropriate point in the angiography procedure.
Selecting
Angiography
will allow the manual capture of FA or ICG images. In Angiography mode, the FA or
ICG timers can be configured to run separately or as a single combined timer. A timer is started by either
selecting the timer or when the initial image captured. The injection should be administered when the scan
head is aligned. If a timer is started accidentally you can reset it by selecting the timer and selecting
Reset
.
You may manually switch between
FA
and
ICG
.
Note
l
You can manually switch between FA and ICG images when both image modalities are being
captured.
l
Auto-capture is available for angiography images, and ensures automatic capture of images when
the scan head is in the correct position relative to the patient.
l
You can select non-angiography image modalities, eye steering and stereo options while performing
an angiography session.
3.2.7.1 Control contrast
You can adjust the contrast for the next image to be captured.
Manually adjust the image contrast by pushing the contrast control to the left or right or by selecting the
desired position on the control.
Depending on the features available on your device you may be able to use the auto contrast. Manually
adjusting the contrast turns auto contrast off. You can switch between
Manual
and
Auto
contrast control
modes.
The contrast is set separately for
opto
map
fa
and
opto
map
icg
images. When you change between
opto
map
fa
and
opto
map
icg
, the device uses the most recent contrast settings for the selected image modalities.
Page 42 of 72
Part Number: G110230/1ENG
English
Copyright 2019, Optos plc. All rights reserved.
Summary of Contents for P200T E Series
Page 1: ...optos com USER GUIDE Optos OCT Ophthalmoscope P200TxE ...
Page 4: ...Page 4 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 10: ...Page 10 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 70: ...Page 70 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 72: ...G110230 1ENG ...