General Use of Smartscope and
Smartscope PRO with M5
Page 1: ...General Use of Smartscope and Smartscope PRO with M5...
Page 2: ...dications for use of Smartscope PRO with M5 Features of M5 device Screen icons indicators Usage Powering on off Attaching optics Taking an image Cradle and USB connection Wifi Battery replacement Prob...
Page 3: ...s imaging and documentation of patient data through digital images and video Supported optics lenses and their intended uses are EY4 Lens for fundus imaging EY3 Lens for fundus imaging ES2 Lens for an...
Page 4: ...FEATURES OF THE DEVICE POWER LED LEFT SOFT KEY ARROW KEYS CONNECTION LED RESET BUTTON RIGHT SOFT KEY MIDDLE KEY...
Page 5: ...device To power off the device with long press Right soft key Open menu with long press Left LED indicator Green Active when device is powered on Blinking when battery is charged Right LED indicator B...
Page 6: ...FEATURES OF THE DEVICE BAYONET CONNECTOR SYSTEM CONNECTOR BAYONET HOLE OPTICS RELEASE BUTTON DUAL ACTION SHUTTER SNAP FOR OPENING BATTERY COVER...
Page 7: ...ear optics DF no lens Patient 1 Battery 100 Time Zoom indicator Diopter compensation indicator in manual focus mode when using EY4 Autofocus Manual focus AF Assist with EY4 Brightness indicator 10 st...
Page 8: ...POWERING ON OFF Camera is powered on by pressing left soft key Camera is powered off by keeping left soft key pressed down over 2 seconds LEFT SOFT KEY...
Page 9: ...ETACHING OPTICS MODULE Optics is attached by placing bayonet legs on the holes and optics is pressed firmly to the device Optics is released by sliding release button BAYONET LEGS BAYONET HOLES OPTICS...
Page 10: ...he dual action shutter to the second position If assist mode is used feature of EY4 optics an image is taken automatically as soon as the aim help circle has turned green Video is captured by keeping...
Page 11: ...Device menu by selecting Edit Patient list User can add Patient ID and Patient name after choosing New patient New patient can be added also after creating new folder Name of the patient is shown in l...
Page 12: ...s possible to select appropriate image viewing program or simply open the folder to view and then store files to the hard disk of the PC WiFi It is also possible to transfer images with WiFi as the ta...
Page 13: ...in the cradle Storing device in the cradle is not harmful for the battery because battery is charged only when charge is dropped below certain limit To connect Smartscope to the cradle gently place i...
Page 14: ...y card provided by Optomed Size of the memory card is 2 4 GB normal SD card 8 GB Wifi card Memory card is placed next to battery Do not detach memory card when device is powered on Device cannot be us...
Page 15: ...t cover by lifting it up Remove old battery by squeezing battery wires with fingers and pulling the connector out from its socket Put the new battery in the same position and attach gently the connect...
Page 16: ...PROBLEM SOLVING In problem situations there are two options to reset the device Original factory settings can be returned by selecting Restore factory defaults which is located in Device menu...
Page 17: ...if behaviour of device is abnormal Reset button is located in a small opening under display Button is marked with a circle Reset button can be pressed with a thin object such as a paper clip Button n...
Page 18: ...If you need support or have any questions please contact Optomed customer service service optomed com...