52
LTDVE4CH-20 | INSTRUCTIONS MANUAL
Figure 14: the Setup pulse generators GN1-GN4 web page
Each pulse generator can be set independently. Each pulse generator is triggered by a selectable
source and can have a different pulse delay and pulse width.
Source
selects the trigger for the generator. It can be chosen between
Off
(default),
Input
TR1
,
Input TR2
,
Input TR3
,
Input TR4
and
Oscillator
. When set to
Off
the generator is
disabled. When set to
Oscillator
the internal free running oscillator is used as the trigger
Pulse delay
is the delay of the generated pulse with respect to the triggering source. The
pulse delay may range from 0 µs to 1,023,000 µs with variable resolution
Pulse width
is the time duration of the generated pulse. The pulse width may range from 1
µs to 1,023,000 µs with variable resolution