DR01 - EN
Version 1 - 2020-04-09
11
Translation of original instruction
DRO1
EN
OP_B
A_DRO
1_GB
.f
m
Step 3. Press
to display the original
parameter value .
(Factory setting=SPEEd 3)
Use the
key to switch between the speed
parameters.
Step 4. When the screen shows the desired parameter,
press the
key and the display shows SPEED.
Step 5. Press
to return to the operation
screen.
1.15
Adaptive adjustment of installation distance
A. Function description: Please run the "Setup" function after each installation to ensure the
measurement accuracy.
Step 1. First place the reading head on the tape, fix the
reading head and the tape (recommendation: about
1mm distance)
Step 2. Press the
button, the
symbol
appears in the upper right corner of the screen to set
the parameter.
Step 3. Press the
button until the DR01
displays "Setup".
Step 4. Press
to confirm your plans.
Step 5. Press
to start the distance
adjustment.
Step 6. Move the reading head at the slowest speed.
(About 14 seconds for 120mm distance)
Distance ( sensor - read head)