BIOLINE
Manual for installation and use
OPTIKON 2000
8 OPERATION
8.1 AUTOTEST SESSION
Turn on the instrument using the switch at the back. The instrument automatically
carries out certain tests, then records the following information on the screen:
•
SELF TEST:
Shows the result of the system test (normally Passed otherwise
the error code ERR.XXX).
•
PRINTER:
Indicates whether the printer is connected (Ready or Absent)
•
PC-LINK:
Indicates whether the connection via RS232 to the personal
computer has been found (Ready or Absent).
One of the sessions indicated in the line at the bottom of the screen can be
selected from this page: A-SCAN - IOL –HELP. A simple command memo can
be printed by pressing the PRINT key or the paper can be pushed forward by
pressing the key “Arrow up” (useful for loading a new roll).
NOTE:
If the self test identifies an error in the memory, a warning message is displayed, before you proceed
to cancel the whole memory:
“DELETE ALL DATA IN MEMORY?: YES”. Press the Enter key. The data on the intraocular lenses
must then be re-entered by the operator.
8.2 A-SCAN MEASURING SESSION
In the A-scan session, the display shows either the parameters for selecting the
operating mode or the results of the individual biometry. In addition the average of a
maximum number of five biometries and the resulting standard deviation is indicated
in large figures.
It is possible to bring up each of the 5 measurements taken and the relating graphs,
by pressing the left and right arrow keys.
The keys lined up under the display allow the most frequently used functions to be
activated.
The A-scan session page provides the following information: