BIOLINE
Manual for installation and use
OPTIKON 2000
Acoustic Output Parameters
according to IEC/EN/DIN 61157
Manufacturer: Optikon 2000 SpA
Device: Bioline (ref.141002)
Transducer: Biometry probe, 10 MHz (ref.143001)
Parameter
Unit Mode
A
Maximum Power
[mW]
0.003 ± 0.001
Peak negative acoustic pressure [
p
-
]
[MPa]
1.3 ± 0.14
Output beam intensity [
I
OB
]
[mW/cm²]
0.006 ± 0.001
Spatial peak temporal average intensity [
I
SPTA
]
[mW/cm²]
0.2 ± 0.04
Focal length [
l
p
] [mm]
22.6
Pulse beam width (6 dB, parallel) [
w
pb6
║
] [mm]
0.75
Pulse beam width (6 dB, orthogonal )[
w
pb6
┴
] [mm]
0.82
Pulse repetition rate [
prr
] Hz
20
Scan repetition rate
[
srr
] Hz
---
[mm] 3.1
Output Beam Dimensions
[mm] 3.4
Arithmetic-mean acoustic working frequency [
f
awf
] [MHz]
10.8
Transducer to transducer output face distance [
l
tt
] [mm]
---
Transducer standoff distance [
l
ts
] [mm]
Contact
Power up fraction
100 %
Acoustical initialization fraction
100 %
Power up mode
Default
Initialization mode
Default
Freeze output
Yes
Modes included
---
Control settings
---
Code 141002EN
5-4
26.10.05 Rev.0