OPTIKON BIOLINE Installation And Operating Manual Download Page 10

BIOLINE 

Manual for installation and use 

OPTIKON 2000 

• 

The standard AC-DC adaptor must be used exclusively to supply the instrument. 

 

Ordinary maintenance 

• 

Do not sterilise the probe in hot air or in a chemical autoclave. 

• 

Do not use solvents or alcohol to clean the exterior of the instrument. 

• 

The probe (and the relating packaging) is not supplied as sterile and must be 
disinfected before it is applied to the cornea. 

 

Precautions relating to the normal use of the instrument 

• 

Gently apply the probe to the moist cornea, avoiding changing its normal 
geometric arrangement. 

• 

If you are in doubt about the accuracy of the measurement carry out a 
calibration test on the probe and in the event of a negative result or incorrect 
functioning, do not use the instrument and seek technical assistance. 

• 

Thermal prints tend to deteriorate in time. For long-lasting storage it is advisable 
to keep a photocopy of the print. 

• 

Turn off the instrument at the switch when not in use

Code 141002EN 

3-2 

      26.10.05  Rev.0 

Summary of Contents for BIOLINE

Page 1: ...5 Rev 0 BIOLINE PORTABLE BIOMETER INSTALLATION AND OPERATING MANUAL OPTIKON 2000 S p A Via del Casale di Settebagni 13 00138 Rome Italy Tel 39 06 8888355 Fax 39 06 8888388 e mail mailto sales optikon...

Page 2: ...A is an ISO 9001 and ISO 13485 certified company which manufactures surgical and diagnostic devices for ophthalmology Its products are manufactured to satisfy the requirements of 93 42 EEC Medical Dev...

Page 3: ...1 8 1 AUTOTEST SESSION 8 1 8 2 A SCAN MEASURING SESSION 8 1 8 3 A SCAN KEY FUNCTIONS 8 2 8 4 SETTING UP THE A SCAN EXAMINATION 8 3 8 5 IOL CALCULATION SESSION 8 5 8 6 FUNCTION OF THE KEYS IN THE IOL...

Page 4: ...THIS PAGE IS INTENTIONALLY BLANK...

Page 5: ...rations repairs are carried out by OPTIKON 2000 S P A authorized personnel the system is used in accordance with its instructions for use the mains electric installation to which the system is connect...

Page 6: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 1 2 26 10 05 Rev 0...

Page 7: ...p A OPTIKON 2000 S P A reserves the right to ascertain in case of faults whether or not the equipment and or its accessories have been altered tampered with in any way or have been damaged by imprope...

Page 8: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 2 2 26 10 05 Rev 0...

Page 9: ...heat or expose it to direct sunlight or high temperatures Do not install the instrument in such a way that the ventilation slots are obstructed The system must never be used in the presence of inflam...

Page 10: ...ied to the cornea Precautions relating to the normal use of the instrument Gently apply the probe to the moist cornea avoiding changing its normal geometric arrangement If you are in doubt about the a...

Page 11: ...ymbols are used on the BIOLINE label Before starting to use the unit you should familiarize yourself with them and their definitions shown in the table SYMBOLS PUBLISHED BY THE IEC SYMBOL DESCRIPTION...

Page 12: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 4 2 26 10 05 Rev 0...

Page 13: ...he distance from the obstacle The measurement of this time and the experimental knowledge of the speed of the echo allow the distance from the object the retina for example from the emitter probe tip...

Page 14: ...ive MDD Technical standards EN 60601 1 EN 60601 1 1 EN 60601 1 2 EN 60601 2 2 ENVIRONMENTAL SPECS Storage temp range 10 C to 70 C humidity 10 100 non condensing Operation temp range 10 C to 40 C humid...

Page 15: ...320 dots thermal Modes of measurement Automatic and manual Measurements in the memory up to 5 with indication of average and standard deviation Echo amplification Manual in 7 levels or Self regulatio...

Page 16: ...nsity ISPTA mW cm 0 2 0 04 Focal length lp mm 22 6 Pulse beam width 6 dB parallel wpb6 mm 0 75 Pulse beam width 6 dB orthogonal wpb6 mm 0 82 Pulse repetition rate prr Hz 20 Scan repetition rate srr Hz...

Page 17: ...e anaesthetic mixture not applicable DIMENSIONS Height 15 cm Width 23 cm Depth 29 cm Weight 2 2 Kg NOTE 1 Weight and dimensions shown are approximate 2 Specifications are subject to change without not...

Page 18: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 5 6 26 10 05 Rev 0...

Page 19: ...This unit has been packaged to minimize the risk of damage in shipment Unwrap the package and examine the components Handle all components with care during unpacking Use caution cutting the wrapping t...

Page 20: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 6 2 26 10 05 Rev 0...

Page 21: ...thanks to backlighting For maximum performance position the instrument facing the observer Adjust the contrast and brightness to the best possible setting 2 CLEAR KEY On the measurement page by press...

Page 22: ...this page all the settings for the lens constants and the biometric examination can be cancelled and the language for printing the user guide can be selected 7 PRINT KEY Sends the printer the results...

Page 23: ...ent plate and the serial number 4 SUPPLY INPUT Input socket low voltage for connection to the feeder model 142001 142002 or to another source with equivalent characteristics 5 MAIN SWITCH Activates an...

Page 24: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 7 4 26 10 05 Rev 0...

Page 25: ...rd by pressing the key Arrow up useful for loading a new roll NOTE If the self test identifies an error in the memory a warning message is displayed before you proceed to cancel the whole memory DELET...

Page 26: ...n small figures to the right of the average The standard deviation can be indicated vocally if the mode PRMT VOX is selected The standard deviation of a series of accurate measurements must not exceed...

Page 27: ...rds to the archiving session if the connection with an external personal computer is present via port RS232 If the connection is not present the autotest page is displayed PRINT Starts up the printing...

Page 28: ...automatic procedure for identifying the type of eye phakic or aphakic is active for the correct calculation of axial length If the A scan is frozen it is possible to measure the axial length relating...

Page 29: ...known The aforementioned constants can be changed and remain stored with the relating lens The result of the calculation is a table with 5 values power and refraction graded in half diopters The power...

Page 30: ...splay the IOL field using the up down arrows 2 Bring up the desired lens right left arrows 3 Press the IOL key to store the first principal lens Repeat the above for the second lens and end by pressin...

Page 31: ...part Use the keys and again to change the decimal figures and end with Enter or move to another parameter requiring a change Prolonged pressure on the keys and produces a rapid increase and decrease...

Page 32: ...21 Omm and 20 9 0 1 20 8 mm Under no circumstances must the temperature of the tube exceed 30 C because in this circumstance the attenuation of the signal does not allow automatic acquisition The rec...

Page 33: ...TE If the equipment is not being used for an extended period the following procedure should be observed 1 Depress the POWER switch to the OFF position to switch off the control console 2 Refer to the...

Page 34: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 8 10 26 10 05 Rev 0...

Page 35: ...Do not use alcohol solvents or abrasives on the control console 9 1 2 ACCESSORIES Treat the accessories as indicated above 9 2 STERILISATION 9 2 1 UNIT The console and footswitch do not need to be ste...

Page 36: ...handled with outmost care and inspected thoroughly after each use for any damage or wear 2 Place the probes in the appropriate holder after each use Avoid propping up the probe on the table if it fal...

Page 37: ...comes out of the upper part You can also push the paper through when the instrument is switched off by rotating the relevant roll fig 3 6 Insert the roll in the appropriate compartment close the panel...

Page 38: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 9 4 26 10 05 Rev 0...

Page 39: ...See chapter 2 3 DURING MEASURING The instrument does not respond to keyboard commands 1 The instrument is awaiting inputs relating to a selected field press Enter 2 The printer is in operation wait un...

Page 40: ...BIOLINE Manual for installation and use OPTIKON 2000 THIS PAGE IS INTENTIONALLY BLANK Code 141002EN 10 2 26 10 05 Rev 0...

Page 41: ...ocket 7 3 Fuses 3 1 G GAIN 8 4 H Holladay Formula 8 5 I Internal lithium battery 8 6 IOL Calculation 8 5 IOL Key 7 2 K K1 8 5 K2 8 5 L LCD Contrast 7 3 Liquid crystal display 7 1 M Main switch 7 3 Mai...

Page 42: ...BIOLINE Manual for installation and use OPTIKON 2000 W Warranty 2 1 Warnings 3 1 WB CATA parameter 8 4 Code 141002EN 11 2 26 10 05 Rev 0...

Reviews: