15
2.4
Calibration sequence
When starting to work with the Pavone, three steps are required to calibrate the system:
In the software these three steps are indicated in the top menu:
Each button opens a menu enabling specific actions. In general, the sequence to follow is from left to
right. In the paragraphs below each step is explained in more detail.
2.5
Find probe: in air
To couple indentation location with imaging data, the reference point of the probe tip is used. Follow the
steps:
1.
To start the procedure, mount the probe, leave plate 1 empty.
2.
Move Y-stage by 30 000 µm so that objective is easily accessible and can move up without
obstructions.
3.
Move probe stage Z down by ~20 000 µm. Move objective stage O up in steps of 500 or 1000 µm.
4.
Rotate a metal screw above the probe stage to loosen up manual XY stage knobs (seeFigure 10,
left image). Using the knobs (see Figure 10, second image from the left), align the probe first by
eye until the cantilever tip is hovering above the center section of the objective (see Figure 10,
images on the right).
5.
Turn on
“
Brightfield
”
and
“
Phase 2
”
and press
“
Auto cam adjust
”
in the software. If the image of
the camera is dark, increase LED intensity
or go to “Camera” settings
. You should see an image or
shadow of the probe. If necessary, adjust the height of the objective stage so that the probe is in
the focal plane (check the working distance of your objective e.g. Nikon WD 3.1 mm for x20
objective, WD 2.1 mm for x40 objective).
6.
Continue alignment of the probe with manual knobs until you see the end of the cantilever as in
Figure 11. Use small steps of the objective stage to get the tip of the probe in focus. Position the
tip at the top of the field of view of the camera.
7.
Tighten back the top screw.
8.
Press
“
Find Probe
”
in software, use the circle tool to mark the circumference of the tip (see Figure
11). Tip: press and hold SHIFT while doing so to keep a perfect circle during resizing. After
Find probe
Wellplates
Calibration
Find probe
Wellplates
Calibration
Click in
probe
Move Y
stage
down
Move Z-
stage
down
Move O-
stage up
Adjust
manually
probe
Focus on
the tip
Find
Probe