- 43 -
optek-Manual--1004-1010-02--ProfibusPA-US-2017-06-23
Specifications
Note!
Enhanced diagnosis information Diag. ID 32 to Diag. ID 34 can be enabled or
disabled per bit mask in the parameter SET_UNCERTAIN_AS_BAD (slot
index 70, physical block). Upon delivery, these bits are disabled.
*
If this diagnosis bit is set, an enhanced error analysis can currently be read
out as an optek-specific error number from the parameter
ERROR_NUMBER (slot index 34, status transducer block).
**
The bit number signals a signal loss at the corresponding measuring result.
*** There is a problem with internal data communication. Please contact us.
Please find our contact details in chapter 11, page 60.
**** If this diagnosis is signaled, the converter may currently be switched off or
booting. If this is not the case, there can be a problem with internal data
communication. Please contact us. Please find our contact details in
32
ERROR NUMBER AVAILABLE*
1
0
optek-internal error code in
Status TB available
33
SIGNAL LOSS MEASUREMENT VALUE 1**
1
34
SIGNAL LOSS MEASUREMENT VALUE 2**
2
35
SIGNAL LOSS MEASUREMENT VALUE 3**
3
36
SIGNAL LOSS MEASUREMENT VALUE 4**
4
37
NO ISB COMMUNICATION***
5
38
NO MODBUS COMMUNICATION****
6
39
7
Tab. 16