- 28 -
optek-Manual--1004-1010-02--ProfibusPA-US-2017-06-23
Specifications
Note!
* REMOTE_CONTROL_STATUS (slot index 64)
With this parameter you can read which remote functionalities can be released
in the C4000 or Haze Control software.
Data type: Unsigned8 as single bits
1 = Corresponding function in converter is released via bus.
0 = Corresponding function in converter is blocked via bus.
0x00000101 = The product change function (via parameter slot index 67) is not
possible via PROFIBUS
®
. The other two remote functions are
released.
** SET_HOLD (slot index 65)
With this parameter, hold can be triggered or canceled in the converter via
PROFIBUS
®
.
Data type: Boolean
0x00
→
false Converter disables an existing system hold state.
0xFF
→
true
Converter enables a system hold state.
24
40
IDENT_NUMBER_SELECTOR
Unsigned8
1
R, W
26
42
FEATURE
DS-68
8
R, W
27
43
COND_STATUS_DIAG
Unsigned8
1
R, W
28
44
DIAG_EVENT_SWITCH
Diag_Event_Switch
50
R, W
36
52
DEVICE_CONFIGURATION
Visible String
32
R
37
53
INIT_STATE
Unsigned8
1
R, W
38
54
DEVICE_STATE
Unsigned8
1
R, W
39
55
GLOBAL_STATUS
Unsigned16
2
R
48
64
REMOTE_CONTROL_STATUS* Unsigned8
1
R
49
65
SET_HOLD**
Boolean
1
R, W
50
66
SET_ZERO_POINT***
Unsigned16
1
R, W
51
67
SET_PRODUCT****
Unsigned16
1
R, W
52
68
CONVERTER_SN
Visible String
16
R
53
69
MODEL_NR
Visible String
16
R
54
70
SET_UNCERTAIN_AS_BAD*****
Unsigned8
1
R, W
Tab. 7
Physical Block (cont.)
Rel. Index
Slot index
Parameter
Data type
Size
(bytes)
Access
BIT 7
BIT 6
BIT 5
BIT 4
BIT 3
BIT 2
BIT 1
BIT 0
not used
hold
product change
system zero point