2.2
pH – C210, C211 and C221 only*
Inputs:
One (C210 and C221) or two (C211) pH or mV inputs and
solution
earth
One (C210 and C221) or two (C211) temperature sensors
Input resistance:
Glass:
>1 x 1013
Ω
Reference:
1 x 1013
Ω
Range:
–2 to 16pH or –1200 to +1200mV
Minimum span:
Any 2pH span or 100mV
Resolution: 0.01pH
Accuracy: 0.01pH
Temperature compensation modes:
Automatic or manual Nernstian compensation
Range –10 to 200°C (14 to 392°F)
Process solution compensation with configurable coefficient
Range 0 to 100°C (32 to 212°F)
adjustable –0.05 to +0.02%/°C (–0.02 to +0.009%/°F)
Temperature sensor:
Programmable Pt100, Pt1000 or Balco 3k
Ω
Calibration Ranges
Check value (zero point):
0 to 14pH
Slope:
Between 40 and 105% (low limit user-configurable)
Electrode Calibration Modes:
Calibration with auto-stability checking
Automatic 1 or 2 point calibration selectable from:
•
ABB
•
DIN
•
Merck
•
NIST
•
US
Tech
2 x user-defined buffer tables for manual entry
2-point calibration or single-point process calibration
* Enables connection to glass or enamel pH and reference sensors and Redox (ORP) sensors
Specifications may be limited by the pH electrode in use. Refer to corresponding