Temperature Value – Point 1
Deg. C (or Deg. F) and Adjust are shown alternately on the upper display line.
Use the and keys to adjust the displayed reading to the required temperature
value for point 1 on the curve.
Ratio Value – Point 1
Ratio and Adjust are shown alternately on the center display line. Use the and
keys to adjust the displayed reading to the required ratio value for point 1 on the
curve.
Temperature and Ratio Values – Points 2 to 6
Enter the temperature and ratio values for the remaining points in the user-
defined temperature compensation curve.
Filter Time
Set the required filter time, between 1 and 60 seconds in 1 second increments.
Sensor B configuration (dual input analyzers only) is identical to
Sensor A configuration.
See Section 5.2.4.