5.1.6 Configure Outputs
Configure Output 1
Output 2 configuration (and Outputs 3 and 4 if option board fitted and
analog features enabled – see Section 6.1.3) is identical to Output 1
configuration.
Assign
Select the sensor and analog output required:
Sen.A
pH/Redox (ORP)/mV for the selected sensor.
Sen.B
Temp.A
Temperature for the selected sensor.
Temp.B
A-B
Difference between the Sensor A and Sensor B readings.
Note.
Sen.B, Temp.B and A-B are applicable only to dual input analyzers.
Range
Select the analog output current range for the selected output.
Span Value
pH (or mV or Deg.C or Deg.F or A-B) and Adjust are shown alternately on the
upper display line. Use the and keys to adjust the displayed reading to the
required span value.
Note.
The minimum and maximum span values are determined by the Zero Value
setting (see next page) plus the minimum differential, e.g. to set Span Value to
0.00 pH, first set Zero Value to –2.00pH.
pH
0.00 to 16.00 pH (minimum differential, 2.00pH)
Redox/ORP
–1100 to 1200 mV (minimum differential, 100mV)
Temperature Deg. C 0.0 to 150.0 (minimum differential, 10°C)
Deg. F 32.0 to 302.0 (minimum differential,18°F)
A-B
–
0.00 to 14.00 pH (minimum differential, 2.00pH)
Note.
A-B is applicable only to dual input analyzers.
Continued on next page.
Config. Output 1
-----
AO1: Assign
-----
AO1: Range
-----
CONFIG. OUTPUTS
-----
A-B
Temp.B
Sen.B
Temp.A
Sen.A
4-20mA
0-20mA
0-10mA
AO1: Zero Value
Config. Output 2
AO1: Span Value
7.00
pH
0.00
pH