5.1.3 Configure Sensors
Configure Sensor A
Sensor B (dual input analyzers only) configuration is identical to
Sensor A configuration.
Probe Type
Select the required probe type.
ORP – Millivolt display
Redox – Millivolt display
pH
– pH display
Differential Input
Yes
– Select if electrode system is equipped with a solution earth (ground) rod
and sensor diagnostics are required – see Section 5.1.4.
No
– Select if electrode system is not equipped with a solution earth (ground)
rod or sensor diagnostics are not required.
Note.
Ensure the electrode system is connected correctly for the type of input
selected (Standard or Differential) – (wall-/pipe-mount analyzers)
(panel-mount analyzers).
Probe Type set to Redox or ORP – continued below.
Probe Type set to pH – continued on next page.
Temperature Sensor
Select the type of temperature sensor used: Pt100, Pt1000, Balco 3K or None.
Enable Calibration
Select Yes to enable sensor calibration.
Select No to disable sensor calibration and all associated frames.
Return to main menu.
See Section 5.1.4.
Config. Sensor A
-----
Config. Sensor B
A: Temp. Sensor
A: Probe Type
-----
ORP
Redox
pH
CONFIG. SENSORS
-----
A: Diff. Input
-----
Yes
No
A: Electrode
A: Temp. Sensor
-----
None
3K Bal
Pt1000
Pt100
A: Enable Cals
-----
Yes
No
A: Probe Type
set to
Redox/ORP
CONFIG. SENSORS
CONFIG. ALARMS
Config. Sensor A