Sensor Slope
Measured conductivity value.
Sensor slope value.
Use the and keys to adjust the sensor slope value within the range 0.200 to
5.000 until the measured conductivity value is correct.
Sensor Offset
Measured conductivity value.
Sensor offset value.
Use the UP and DOWN keys to adjust the sensor offset value within the range –
20.00 to 20.00 until the measured conductivity value is correct.
Temperature Slope
Measured temperature value.
Temperature slope value.
Use the UP and DOWN keys to adjust the temperature slope value within the
range 0.200 to 1.500 until the measured temperature value is correct.
Temperature Offset
Measured temperature value.
Temperature offset value.
Use the and keys to adjust the temperature offset value within the range –40.0
to 40.0°C (–40.0 to 104.0°F) until the measured temperature value is correct.
Sensor B calibration (dual input analyzers only) is identical to
Sensor A calibration.
Single input analyzers only – return to top of page.
Reset Calibration
Select Yes and press
DOWNSCROLL
to reset the calibration data.
Select No and press
DOWNSCROLL
to abort.
Return to top of page.