2.1.2 Operating
page
2.1.2.1 Single Input pH
Measured Values
pH.
Temperature.
Measured Millivolts
Millivolts.
% Slope and pH Check Value
% slope value.
A value between the programmed minimum % slope value (see
Set Min Slope
in the
CONFIG. SENSORS
page – Section 5) and 105% is displayed. If the value is outside
these limits, check the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7pH is the
optimum value for glass electrodes and 0pH for Antimony electrodes.
Alarm 3 set to
Wash
(
A3: Type
, Section 5) – see Section 2.1.2.6
Alarm 3 not set to
Wash
.
Monitoring pH
7.00
pH
24.4
Deg.C
Wash Function
Monitoring pH
Millivolts
404
mV
A: Slope & Check
100.0
%
7.00
pH