Factory Settings Access Code
Enter the required code number, between 0000 and 19999, to gain access to the
factory settings. If an incorrect value is entered, access to subsequent frames is
prevented and the display reverts to the top of the page.
Calibrate Sensor A
Note.
The values in the display lines for sensor calibration are shown as
examples only – the actual values obtained will differ.
Sensor B calibration (dual input analyzers only) is identical to
Sensor A calibration.
Single input analyzers only – see page 97.
Operating Page – see Section 2.2.2.
Calibrate Input for Sensor A ?
If calibration is required select Yes otherwise select No.
Note.
To abort calibration, press the
DOWNSCROLL
key again at any time
before calibration is complete – see next page.
Resistance Span (25K
Ω
)
Set the cell simulator to 25k
Ω
.
The display advances automatically to the next step once a stable and valid
value is recorded.
Note.
The upper 6-segment display shows the measured voltage. Once the
signal is within range the lower 6-segment display shows the same value and
Calib is displayed to indicate that calibration is in progress.
Resistance Span (2.5K
Ω
)
Set the cell simulator to 2k
Ω
.
The display advances automatically to the next step once a stable and valid
value is recorded.
Resistance Span (250
Ω
)
Set the cell simulator to 250
Ω
.
The display advances automatically to the next step once a stable and valid
value is recorded.
Continued on next page.
Factory Set Code
0000
FACTORY SETTINGS
-----
Cal. Sensor A
-----
A: Res 2.5 K ohm
2.500
2.500
A: Res 250 ohm
250.0
250.0
A: Res 25 ohm
Cal. Sensor B
Cal. Output 1
Conductivity
A: Res 25 K ohm
25.00
25.00
Kohms
Calib
Calibrate I/P ?
-----
Yes
No
Yes
No
Kohms
Calib
Ohms
Calib
Dual Cond.