Calibrate Output 1
Note.
When adjusting the 4 and 20mA outputs, the display reading is
unimportant and is used only to indicate that the output is changing when the
UP
and
DOWN
keys are pressed.
See below.
Adjust 4mA
Use the
UP
and
DOWN
keys to set the milliammeter reading to 4mA.
Note.
The analog output range selected in Configure Outputs (see Section 5.1.6)
does not affect the reading.
Adjust 20mA
Use the and keys to set the milliammeter reading to 20mA.
Note.
The analog output range selected in Configure Outputs (see Section 5.6)
does not affect the reading.
See below.
Operating Page – see Section 2.1.2.
Calibrate Output 2
Note.
Output 2 calibration is identical to Output 1 calibration.
Option board fitted and analog features enabled – see next page.
Option board fitted, additional features disabled – see next page.
Option board not fitted – see next page.
Option board fitted and analog features enabled – see next page.
Option board fitted, additional features disabled – see next page.
Option board not fitted – see next page.
Operating Page – see Section 2.1.2.
Cal. Output 1
-----
O1: Adjust 4mA
16000
O1: Adjust 20mA
7200
Monitoring pH
Cal. Output 2
Cal. Output 1
Cal. Output 2
Cal. Output 2
-----
O1: Adjust 20mA
7200
Monitoring pH
Cal. Output 3
Cal. Output 2
Cal. Output 3
Option Board
Option Board
Alter Fact. Code
Alter Fact. Code
Monitoring Redox
Dual Redox
pH/Redox
Monitoring Redox
Dual Redox
pH/Redox