6.1.2 Preparation
a) Switch off the supply and disconnect the electrode system, temperature compensator(s) and current output(s) from
the analyzer's terminal blocks.
b) Sensor A:
1) Link terminals B9 and B10.
2) Connect the millivolt source to terminals B14 (–ve) and B16 (+ve) to simulate the pH or Redox input. Connect the
millivolt source earth to the Case Earth (Ground) Stud – see Manual CONTROL 200 Part 1 - Hardware.
3) Connect the 0 to 10k
Ω
decade resistance box to terminals B9 and B11 to simulate the Pt100/Pt1000/Balco 3K.
Sensor
B:
1) Link terminals B1 and B2.
2) Connect the millivolt source to terminals B6 (–ve) and B8 (+ve) to simulate the pH or Redox input. Connect the
millivolt source earth to the Case Earth (Ground) Stud – see Manual CONTROL 200 Part 1 - Hardware.
3) Connect the 0 to 10k
Ω
decade resistance box to terminals B1 and B3 to simulate the Pt100/Pt1000/Balco 3K.
c) Connect the milliammeter to the analog output terminals.
d) Switch on the supply and allow ten minutes for the circuits to stabilize.
d) Select the FACTORY SETTINGS page and carry out Section 6.1.3.