LBL-2019-41-v4 OptoMonitor 3 IFU wTAVI_en Rev 2022-03
55
|
P a g e
5.13
Input/Output Calibration with Cathlab Hemodynamic System, and
Time/Date/Language Adjustments via the Maintenance Menu
5.13.1
Accessing the Maintenance Menu
To access the Maintenance menu, begin by entering the SETTINGS menu (0).
1-
Press the Maintenance button. This will take you to the Maintenance menu.
2-
Enter the maintenance access code and
press the green √ to continue.
Refer to section 8.12 for
default values.
3-
Select Input/Output Calibration
4-
Input/Output Calibration screen will be shown: