5.2
TAVI/TAVR Mode
This setting is only available if the software has TAVI capabilities. Check your activation key in Activation
Management Menu to see what software features are available on your device.
When TAVI/TAVR is enabled, the signal labels, colors and averaging will change. These settings are
independent from the settings available when TAVI/TAVR is disabled. The average setting default value is
5 and is saved in non-volatile memory.
5.3
Prospective Equalization
This setting is only available if the software has FFR/dPR capabilities. Check your activation key in
Activation Management Menu to see what software features are available on your device. This setting
only applies to FFR/dPR recordings.
Prospective Equalization can be enabled or disabled. By default, it is disabled. Whether prospective
equalization is selected or not, the equalization calculation takes place over 4 beats.
If “Prospective Equalization” is disabled, then the 4 beats before the equalization button is pressed are
used to perform the equalization. Therefore, equalization seems instant to the operator as there is no
delay between the moment the operator presses on the equalization button and the moment the
equalization is done. However, the equalization button is only available once 4 beats have been recorded.