LBL-2019-41-v4 OptoMonitor 3 IFU wTAVI_en Rev 2022-03
41
|
P a g e
•
One Display Unit with DU power supply, OU-DU communication cable, mounting clamp and
screws.
Depending on intended Cathlab-OptoMonitor 3 connection, the box
may
also include any of the following
cables:
•
Interface cable for Pa input (from the hemodynamic system or Aortic Transducer)
•
Interface cable for Pd output to the hemodynamic system
•
Interface cable for Pa output to the hemodynamic system
•
One DICOM cable – optional
NOTE:
Before use, always check that all parts are present and undamaged.
4.2
Mounting the devices
4.2.1
Mounting the devices
There are several possibilities for mounting the different parts of the OPM3. Please refer to part
Appendix A: OptoMonitor 3 Installation Instructions
for further installation instructions.
4.2.2
Mounting the Optical Unit
The Optical Unit is suitable for installation within patient environment. Install OptoMonitor 3 Optical Unit
securely under the bed or other nearby location. Ensure the Optical Unit and Display Unit are close enough
to connect the OU-DU communication cable, or if using wireless connection, ensure there is no
interference that would interrupt the signal between the OU and DU.
NOTE
: If replacing a OptoMonitor, ensure power supply is replaced as well. OptoMonitor power supply is
not authorized for use with the OpM3.
4.2.3
Mounting the Display Unit
The Display Unit is suitable for installation within patient environment. If using the hardwired connection
between OU and DU, follow these steps to mount the OptoMonitor 3 Display Unit bedside:
1-
Attach the clamp and adapter plate to the rear of the OptoMonitor 3 Display Unit with the
supplied screws.
2-
Affix OptoMonitor 3 Display Unit to a suitable place on an IV pole (not shown). Ensure that the
mounting of OptoMonitor 3 Display Unit is secured and that its position is stable.
3-
Connect the
OU-DU communication cable
(Blue COM wire to COM1 in DU-10”, LAN cable to LAN2 in DU-15”).
It is also possible to use a third-party screen support if needed. The DU-10” and the DU-15” can both be
attached to a VESA screen support. The DU-10” can be attached to a 75x75 VESA support and the DU-15”
can be attached to a 10x10 VESA support.
If using the wireless connection between OU and DU, install the DU in a location which is accessible and
safe to use, and can maintain consistent wireless connection with the OU. This includes any acceptable
location within the control room. Wireless dongle can be connected in any free USB port, but should not
be connected using an extension cable.