LBL-2019-41-v4 OptoMonitor 3 IFU wTAVI_en Rev 2022-03
118
|
P a g e
Assembly 8
Step 1
Step 2
Step 3
Installing the
OpM3 DU-
15” on the
desk stand
Instructions
Attach the plate to the back
of the screen.
Screw the plate to the base
Final view
16
Parameters
Setting parameters shall be settled as follow, unless instructed otherwise by in-hospital personnel:
Averaging:
3
beats
Pressure scale:
0 - 150
mmHg or
0 - 200
mmHg
Time scale:
6
mm/s
Record: Proper parameterization is confirmed by initialing the Appendix B
17
AUX In Calibration Using the OptoMonitor 3 and Cathlab
Hemodynamic System
1.
Navigate to the Maintenance screen via the SETTINGS screen. Press “Input/Output
Calibration” button located in the middle of the top row of the screen. Ensure “Pa Source”
is set to “Aux”.
2.
In the “Pd Out” box, click the “Output (mmHg)” drop-down menu and select “0”.
3.
Zero Cathlab BP1 channel.
4.
Ensure that “Reading (mmHg)” in the “AUX In” box shows “0.0”. If it does
not
, adjust the
offset values until “Reading (mmHg)” shows “0.0”. When this occurs, proceed to step 5.
5.
Next, in the “Pd Out” box, click the “Output (mmHg)” drop-down menu and select “200”.
Check the Cathlab screen and ensure that Cathlab Pa is between 180-220 mmHg.
6.
In the “AUX In” box, ensure that “Reading (mmHg)” shows a value +/-1 from what is being
shown on the Cathlab screen. (e.g., if Cathlab reads 196 mmHg, “OptoMonitor 3 shall read
Cable Connection
1.
Pd output cable connects to BP1 Hemo channel; BP1 is the patient aortic transducer
pressure input channel
2.
Pa input cable connects to Hemo high level output port. High level output is 100mmHg/V