POWER-UP SEQUENCE
The camera goes through the following power-up sequence:
1. A black screen, followed by the Opgal logo, are displayed for
approximately ten seconds.
2.
The cooler starts.
3. The camera starts in
4. The
cooler
icon (A) blinks for 8-10 minutes, indicating the camera is
in cool down state.
5. The camera performs an automatic NUC when cool down is
complete (see
Performing Non-Uniformity Correction (NUC)
).
6. Once cool down is complete, all viewing modes become available
and cooler icon (A) disappears.
Replace this image
NOTES
▪
The cool down process allows the infrared detector to reach operating
cryogenic temperature. Once this temperature is achieved, the cooler
icon disappears, indicating cool down is complete.
▪
During the cool down process the camera menu is available (see
).
A
Operation | Camera Startup
25.
MENU
EyeCGas® 2.0 | User Manual Rev 01 | Doc P/N: UM-8G9A0000B