C h a p t e r 9 P R O C E S S R E C I P E
7/26/00
INTEGRA Operations Revision A
9-41
20-0200-231 (CR), 20-0200-232 (Std), 20-0200-251 (CD)
.
Change IDLE CHEM
33 Toggle the IDLE CHEM ON or OFF (default: OFF) for Brush
Station #1 or Brush Station #2. The Idle Chem icon only displays
when Lo Toxic chemicals such as NH
4
OH, DI, etc. are used. The
Idle Chem icon does
not
display when Hi Toxic chemicals such as
HF, MCC, etc. are used.
Once the system is in IDLE, and the idle time elapses (BRUSH
IDLE TIMER), the IDLE CHEM will turn ON to clean the
brushes. Lo Toxic Brush cleaning continues for the interval time
set in the BRUSH IDLE ON TIME. Brush cleaning stops for the
interval time set in the BRUSH IDLE OFF TIME.
If the
IDLE CHEM,
and/or the
IDLE LOW FLOW DI
, and/or the
IDLE HI FLOW DI
are all set to
ON
, then they will all start at the
same time.
STANDARD