Battery Information 

The logger requires one user-replaceable CR2032 3V lithium 
battery. Battery life is 1 year typical at 25°C (77°F)

 

with a 

logging interval of 1 minute and Bluetooth Always On enabled 
or 2 years typical at 25°C (77°F)

 

when the logger is configured 

with Bluetooth Always On disabled. Expected battery life varies 
based on the ambient temperature where the logger is 
deployed, the logging interval, the frequency of connections, 
downloads, and paging, and the use of burst mode or statistics 
logging. Deployments in extremely cold or hot temperatures or 
a logging interval faster than one minute can impact battery 
life. Estimates are not guaranteed due to uncertainties in initial 
battery conditions and operating environment. 

To replace the battery: 

1.

 

While pushing down with both thumbs on the back of the 
logger, rotate the battery cover counterclockwise until it 
stops moving (about a 1/8 turn). 

 

2.

 

Use the edge below the arrow on the cover to lift it off the 
logger. 

 

3.

 

Remove the battery and place a new one in the battery 
holder, positive side facing up. Use a small flat-head 
screwdriver to carefully pop the battery out of its holder as 
shown. 

 

4.

 

Inspect the O-ring on the battery cover. Make sure it is 
clean and seated properly. Remove any dirt, lint, hair, or 
debris from the O-ring. If the O-ring has any cracks or tears, 
replace it as follows: 

a.

 

Spread a small dot of silicone-based grease on the O-
ring with your fingers, making sure the entire O-ring 
surface is completely covered in grease. 

b.

 

Place the O-ring on the cover and clean off any debris. 
Make sure the O-ring is fully seated and level in the 
groove and not pinched or twisted. This is necessary to 
maintain a waterproof seal. 

5.

 

Place the cover back on the logger as shown, aligning the 
notches in the cover with the rounded tabs on the logger 
case. The cover does not close properly and maintain a 
waterproof seal if it is misaligned. 

 

6.

 

While pushing down with both thumbs, rotate the battery 
cover clockwise until it locks in position under the two large 
corner tabs and the two small rounded tabs. 

 

 

WARNING:

 Do not cut open, incinerate, heat above 85°C 

(185°F), or recharge the lithium battery. The battery may 
explode if the logger is exposed to extreme heat or conditions 
that could damage or destroy the battery case. Do not dispose 
of the logger or battery in fire. Do not expose the contents of 
the battery to water. Dispose of the battery according to local 
regulations for lithium batteries.

 

 

Use this edge to 
lift the cover off 

Align notch in cover with 
small, rounded tab on logger 

Cover secured under 
rounded tab 

Cover secured 
under corner tab 

Cover secured 
under corner tab 

Cover secured under 
rounded tab 

Align notch in cover with 
small, rounded tab on logger 

Summary of Contents for HOBO Pendant MX Temp

Page 1: ...30 5 m 100 ft Radio Power 1 mW 0 dBm Transmission Range Approximately 30 5 m 100 ft line of sight Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 second to 18 hours Time Acc...

Page 2: ...ging Press the button for 10 seconds to reset a password Note The circle on the front of the logger represents the button area on the logger You will not feel an actual button push when you press that...

Page 3: ...the firmware on the logger A logger readout is completed automatically at the beginning of the firmware update process Important Before updating the firmware on the logger check the remaining battery...

Page 4: ...rm value 8 For the Duration select how much time should elapse before the alarm trips and select one of the following Cumulative Samples The alarm trips once the sensor reading is out of the acceptabl...

Page 5: ...logging mode a New Interval event is automatically logged and the burst condition is cleared even if the actual high or low condition has not cleared Statistics Logging During fixed logging the logge...

Page 6: ...load data automatically to HOBOlink Onset s web based software using the app or the MX gateway For details see the HOBOconnect User Guide and see the HOBOlink help for details on working with data in...

Page 7: ...Guide at www onsetcomp com manuals rs1 Note To log both temperature and sunlight you need two MX2202 loggers Place one logger in a solar radiation shield to log temperature and the other mounted flat...

Page 8: ...ver Make sure it is clean and seated properly Remove any dirt lint hair or debris from the O ring If the O ring has any cracks or tears replace it as follows a Spread a small dot of silicone based gre...

Page 9: ...tion between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technicia...

Reviews: