background image

Pulse Input Adapters (Part # S-UCC/D-M006) 

for use with HOBO

®

 H21 and H22 Series Data Loggers

 

 

© 2007 Onset Computer Corporation 

The Pulse Input Adapters are used to log the number of switch 
closures per interval and are designed to work with smart 
sensor-compatible HOBO data loggers. The adapter has a 
plug-in modular connector that allows it to be added easily to 
these loggers. Two versions are available: one for mechanical 
contact closures (ideal for connecting tipping-bucket rain 
gauges to HOBO H21 series loggers), and one for solid-state 
electronic switches (used for connecting compatible pulse 
output sensors to HOBO H22 series loggers). 

Inside this package 

 

Pulse Input Adapter 

 

2 wire nuts 

 

Pulse Input Adapter 

Specification 

S-UCC-M006 

for Electronic Switches 

S-UCD-M006 

for Contact Closures 

Maximum Input Frequency 

120 Hz (120 pulses per second) 

2 Hz (2 pulses per second) 

Measurement Range

 

0 – 65,533 pulses per logging interval 

Resolution

 

1 pulse 

Lockout time 

45 µs ± 10% 

327 ms ± 10% 

Recommended Input Type 

Electronic solid state switch closure or 
CMOS-level digital output (example: 
FET, opto-FET or open collector) 

Mechanical contact closure (example: 
reed switch in a tipping-bucket rain 
gauge) 

Preferred Switch State

 

Active low input 

Normally open  

Edge Detection 

Falling edge, Schmitt Trigger buffer (logic levels: low 

 0.6 V, high 

 2.7 V) 

Minimum Pulse Width 

1 ms 

Input/Output Impedance 

100 K

 

Open Circuit Input Voltage 

3.3 V 

Maximum Input Voltage 

3.6 V 

User Connection 

24 AWG wires, 2 leads: white(+), black(-) 

Operating Temperature 
Range 

-40° to 75°C (-40° to 167°F) 

Overall Cable Length 

6.5 m (21 ft.) 

Housing 

Weatherproof PVC housing protects input adapter electronics 

Housing Dimensions 

14 x 0.95 cm (5.5 x 0.375 in.) 

Weight 

310 g (11 oz.) 

Bits per Sample

 

16  

Number of Data Channels 

Measurement Averaging 
Option 

No (reports the number of pulses over the logging interval) 

Part Number

 

S-UCC-M006 S-UCD-M006 

 

   For maximum battery life, the Pulse Input Adapters should be used with their preferred switch type. The adapters 

will work with active high inputs (S-UCC) and normally closed switches (S-UCD), but battery life will not be 
optimized. 

Part #: MAN-S-UCC/D 