FT-3100 TRANSMITTER PROGRAMMING GUIDE
ONICON Incorporated 727.447.6140
Page 10
onicon.com
Select the sensor body liner material from the following list.
Liner Material*
Code
PTFE
PTFE
Polypropylene
PP
Ebonite
HR
*Disregard all liner material options that are not shown in this list.
To change the setting, press enter to access the function. Use the right arrow key
to move the cursor and the up arrow key to change the code as per the table
above. Press enter to accept the change and exit the function. Press escape to
This parameter configures the transmitter for U.S. (imperial) or SI (metric) units.
To change the existing setting, press enter to access the function and use the up
arrow key to change the setting. Press enter to accept the change and exit the
function. Press escape to exit the program mode.
This must be set to match the nominal diameter of the sensor body connected
to the transmitter. If the Unit Type is set to U.S. units, the nominal sizes will be in
inches. For SI units, the nominal sizes will be in millimeters.
To enter a new diameter, press enter to access the function. Use the right arrow
key to move the cursor and the up arrow key to change the value. Press enter to
accept changes and exit the function. Press escape to exit the program mode.
KA is one of two coefficients of calibration that are derived during the sensor
body factory calibration. Both variables must be programmed in to the
transmitter to ensure accurate flow measurement. The KA value is provided on
the certificate of calibration and on a label on the sensor body.
To enter a new coefficient, press enter to access the function. Use the right arrow
key to move the cursor and the up arrow key to change the value. Press enter to
accept changes and exit the function. Press escape to exit the program mode.
KZ is the second of two coefficients of calibration that are derived during the
sensor body factory calibration. Both variables must be programmed in to the
transmitter to ensure accurate flow measurement. The KZ value is provided on
the certificate of calibration and on a label on the sensor body.
To enter a new coefficient, press enter to access the function. Use the right arrow
key to move the cursor and the up arrow key to change the value. Press enter to
accept changes and exit the function. Press escape to exit the program mode.
Liner Material
[Lining = XXXXXX]
Unit Type
[U.type = XXXXXX]
Diameter
[Diam. = in XXX]
KA
[KA = ± XX.XXX]
KZ
[KZ = ± XXXX]