10. POSTOPERATIVE INFORMATION
10.1 Magnetic Resonance Imaging (MRI) Compatibility
MR Conditional
The
On-X Prosthetic Heart Valve, Conform-X Mitral Heart Valve
Prosthesis, Size 25-33*
, was determined to be MR-conditional accord-
ing to the terminology specified in the American Society for Testing and
Materials (ASTM) International, Designation: F2503-08. Standard Practice
for Marking Medical Devices and Other Items for Safety in the Magnetic
Resonance Environment. ASTM International, 100 Barr Harbor Drive, PO
Box C700, West Conshohocken, Pennsylvania.
Non-clinical testing demonstrated that the On-X Conform-X Mitral Heart
Valve Prosthesis, Size 25-33, is MR Conditional. A patient with this device
can be scanned safely immediately after placement under the following
conditions:
Static Magnetic Field
• Static magnetic field of 3-Tesla or less
• Maximum spatial gradient magnetic field of 720-Gauss/cm or less
MRI-Related Heating
In non-clinical testing, the On-X Prosthetic Heart Valve, Conform-X Mitral
Heart Valve Prosthesis, Size 25-33, produced the following temperature
rise during MRI performed for 15-min of scanning (i.e., per pulse sequence)
in the 3-Tesla (3-Tesla/128-MHz, Excite, HDx, Software 14X.M5, General
Electric Healthcare, Milwaukee, WI) MR system:
Highest temperature change +1.6˚C
Therefore, the MRI-related heating experiments for the On-X Conform-X
Mitral Heart Valve Prosthesis, Size 25-33, at 3-Tesla, using a transmit/receive
RF body coil at an MR system reported whole body averaged SAR of 2.9-W/
kg (i.e., associated with a calorimetry measured whole body averaged value
of 2.7-W/kg), indicated that the greatest amount of heating that occurred in
association with these specific conditions was equal to or less than +1.6˚C.
Artifact Information
MR image quality may be compromised if the area of interest is in the exact
same area or relatively close to the position of the On-X Conform-X Mitral
Heart Valve Prosthesis, Size 25-33. Therefore, optimization of MR imaging
parameters to compensate for the presence of this device may be neces-
sary.
Pulse Sequence
T1-SE
T1-SE
GRE
GRE
Signal Void Size
1,090-mm
2
686-mm
2
1,478-mm
2
1,014-mm
2
Plane Orientation
Parallel
Perpendicular
Parallel
Perpendicular
*The MRI findings apply to this specific largest heart valve prosthesis and all
other smaller sizes which are made from similar materials.
Summary of Contents for Ascending Aortic Prosthesis
Page 23: ...On X 6 in situ On X 5 93 42 I 13 6h AAP 6...
Page 29: ...On X 12 3 8 9 6 19 Gelweave Valsalva Vascular Prosthesis 5 9 7 On X LTI...
Page 32: ...On X 15 12 On X LTI On X LTI On X LTI On X LTI On X LTI On X LTI On XLTI...
Page 152: ...On X 2 442 1 On X 442 2 On X 442 3 443 4 443 5 443 6 443 7 443 8 444 1 On X mm 445 2 446...
Page 155: ...On X 5 On X On X LTI 19 mm 25 mm 27 29 mm 4 2 On X On X 19 Gelweave Valsalva 5 On X 5...
Page 156: ...On X 6 93 42 I 13 6 AAP 6 6 1 On X 7...
Page 161: ...On X 11 19 mm 25 mm 19 mm 25 mm 19 mm 25 mm 27 29 mm 9 5 3 8 9 6 19 GelweaveValsalva 5...
Page 164: ...On X 14 12 On X LTI On X LTI On X LTI On X LTI On X LTI On X LTI On X LTI...
Page 199: ...On X Ascending Aortic Prosthesis IFU 6 Gelweave Valsalva 5 On X 5 93 42 EEC I 13 6h AAP...
Page 210: ...On X IFU 2 442 1 On X 442 2 On X 442 3 443 4 443 5 443 6 443 7 443 8 444 1 On X mm 445 2 446...
Page 214: ...On X IFU 6 On X 5 13 6h I 93 42 EEC AAP 6 6 1 On X...
Page 220: ...On X IFU 12 3 8 9 6 19 Gelweave Valsalva 5 9 7 On X LTI...
Page 271: ...On X IFU 5 On X On X On X On X LTI On X On X LTI On X On X LTI 25 19 27 29 4 2 On X On X 19...
Page 272: ...IFU On X 6 Gelweave Valsalva 5 On X 5 93 42 EEC 13 6h AAP...
Page 273: ...On X IFU 7 6 6 1 On X 7 8 8 1 SZ 27 29 25 23 21 19 On X mm 1 8 2 On X 2...
Page 278: ...IFU On X 12 25 19 25 19 25 19 27 29 9 5 3 8 9 6 19 Gelweave Valsalva 5...
Page 281: ...On X IFU 15 On X LTI 11 2 12 On X LTI On X LTI On X LTI On X LTI On X LTI On X LTI...
Page 327: ...On X 2 442 1 On X 442 2 On X 442 3 443 4 443 5 443 6 443 7 443 8 444 1 On X 445 2 446...
Page 331: ...On X 6 5 93 42 EEC I 13 6h AAP 6 6 1 On X...
Page 332: ...On X 7 7 8 8 1 On X 19 21 23 25 27 29 SZ mm 1 8 2 On X 2 8 3 On X Tyvek On XLTI 2 1 2 On X...
Page 336: ...On X 11 12 9 4 On X On X 19 25 27 29 19 25 19 25 19 25 27 29 9 5 3 8 9 6...
Page 337: ...On X 12 19 GelweaveValsalva 5 9 7 On X LTI On X 9 8 On X On X...
Page 341: ...On X 2 442 1 On X 442 2 On X 442 3 443 4 443 5 443 6 443 7 443 8 444 1 On X mm 445 2 446...
Page 345: ...On X 6 On X 5 93 42 I 13 6 AAP 6 6 1 On X...
Page 346: ...On X 7 7 8 8 1 On X 19 21 23 25 27 29 mm SZ mm 1 8 2 On X 2 8 3 On X Tyvek On X LTI 2 1 2...
Page 351: ...On X 12 3 8 9 6 19 Gelweave Valsalva 5 9 7 On X...
Page 354: ...On X 15 12 On X LTI On X LTI On X LTI On X LTI On X LTI On X LTI On X LTI...
Page 430: ...On X 2 442 1 On X 442 2 On X 442 3 443 4 443 5 443 6 443 7 443 8 444 1 On X 445 2 446...
Page 433: ...On X 5 19 25 27 29 4 2 On X On X 19 C Gelweave Valsalva 5 in situ On X 5...
Page 434: ...On X 6 93 42 EEC I 13 6h 6 6 1 On X 7 8 8 1 On X 19 21 23 25 27 29 SZ mm 1 8 2 On X 2...
Page 439: ...On X 11 9 6 19 C Gelweave Valsalva 5 9 7 On X in situ in situ On X 9 8 On X On X...
Page 444: ...On X Ascending Aortic Prosthesis IFU Figure 8 Removing Holder Handle from Prosthesis 444...