Setting the Output Conditions
ZW User's Manual
93
4
Settings for
Func
tion
Setting the Differentiation Filter
Set this filter to detect sudden changes in measured values occurring within an extremely short time.
The differentiation filter detects changes in the measured value and current value before a comparison interval.
The time of this comparison interval is defined as the differential cycle. (Default value: OFF)
Explorer pane
: [Bank] | [(Bank Data Name)] (double click)
Edit pane
: [Task Settings] icon (
)
Task Settings window : [Filter]
Item
Setting item
Setting value
Description
Differentiation
filter
Filter mode
OFF (default value)/ON
Set the differentiation filter ON/OFF.
Differentiation cycle
1 (default value) to
5000 [ms]
Set the internal time for the measurement value to be
compared with the current value.
1
Select differentiation filter ON/OFF from
[Differentiation filter] - [Filter mode].
2
Set the number of differential cycles from
[Differentiation cycle].
The differentiation filter can also be set by the operating keys on the Sensor Controller.
Setting the Differentiation Filter p.239
Display value
(differentiation result)
Sampling
Present value
Comparison interval: 10 ms
Measuring cycle
Note
Summary of Contents for ZW-C1*T
Page 70: ...Operating with Sensor Controller 68 ZW User s Manual MEMO ...
Page 116: ...Setting the System 114 ZW User s Manual MEMO ...
Page 134: ...Controll input signal with PC tool 132 ZW User s Manual MEMO ...
Page 209: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 208 8 2 Troubleshooting 210 ...
Page 319: ...Index ZW User s Manual 317 Index 11 ...
Page 321: ......