Setting the Output Conditions
ZW User's Manual
95
4
Settings for
Func
tion
Setting the Average Filter
Output measured data as the average value of a preset count.
Set this, for example, to ignore sudden changes in measured value.
Explorer pane
: [Bank] | [(Bank Data Name)] (double click)
Edit pane
: [Task Settings] icon (
)
Task Settings window : [Filter]
When measurement is started, the measurement result is not updated and output until the measured values have
been sampled for the preset average count.
Item
Setting item
Setting value
Description
Average number
of times
Average
1/2/4/8/16/32/64/128/
256 (default value)/512/
1024/2048/4096 [times]
Switches the average count.
1
Select how many times the latest measured
values are to be averaged before output from
[Average number of times] - [Average].
The average filter can also be set by the operating keys on the Sensor Controller.
Setting the Average Filter p.311
Important
Note