![Omron ZW-7000 Series User Manual Download Page 156](http://html1.mh-extra.com/html/omron/zw-7000-series/zw-7000-series_user-manual_742279156.webp)
Setting the Output Conditions
154
ZW-8000/7000/5000
User's Manual
Setting the Average Filter
Output measured data as the average value of a preset count.
This is set in cases such as to reduce the fluctuation of measurement values and increase the static resolution.
Multi View Explore
: [Bank Group] | [(Bank Data Name)] (double click)
→
Edit pane
: [Task Settings] icon (
)
→
Task Settings window : [Filter]
When measurement is started, the measurement result is not updated and output until the measured values have
been sampled for the preset average count.
Item
Setting item
Setting value
Description
Average number
of times
Average
• Internal synchronous measurement mode
1/2/4/8/16/32/64/128/256/512/1024(default value)/2048/4096/
8192/16384 [times]
Switches the average count.
• External/PDO synchronous measurement mode
1(default value)/2/4/8/16/32/64/128/256/512/1024/2048/4096/
8192/16384 [times]
1
Select how many times the latest measured
values are to be averaged before output from
[Average].
The average filter can also be set by the operating keys on the Sensor Controller.
Setting the Average Filter p.243
Important
Note
Summary of Contents for ZW-7000 Series
Page 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Page 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Page 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Page 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Page 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Page 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Page 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Page 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Page 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Page 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Page 335: ......